DD+DIS150.06E
8.1
New Features of Software CRE_1509
8.1.1
New features for the Customer
Some minor improvements in IP handling and communication between digitizer
and NX processing station. Details see section 8.2.4.
8.1.2
New features for Service
Software CRE_1509 supports both, the existing PMI board CM+9 5175 2290 0*
and the new developed PMI2 board CM+9 5175 2180 0 (which is not compatible
to software < CRE_1509).
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
Result of stall calibration is shown in stall calibration user interface as well as
stored in the service activity log file. This result can also be used for
troubleshooting. For details refer to section 3.2, Tools.
Software CRE_1509 displays error 4642 after replacement of a spare part with
E-label (i.e. each spare part which keeps spare part specific data):
This error can only be solved in service client menu "Configuration – Exchanged
Components". There you have to enter, whether the spare part is a complete
module, or whether it is a printed circuit board only.
Example:
- For replacement of slow scan board select "Board".
- For replacement of "slow scan unit" (incl. slow scan board) select "Module".
This new feature was required due to logistical reasons. For details refer to
Technical Documentation, chapter 3.2, Tools.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
Figure 1
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
Agfa Company Confidential
Chapter 3.8 / 13