DD+DIS150.06E
DISASSEMBLY:
(1)
Switch off the digitizer.
(2)
Disconnect the mains cable.
(3)
Simultaneously press the two buttons (A)
positioned underneath the feed table.
(4)
Pull out cassette unit (B).
(5)
With a screwdriver remove required
bearing at right hand side (see Figure 101)
or left hand side (see Figure 99).
See NOTE below.
NOTE:
As the ball bearings are secured with Loctite, some force has to be applied to remove the
ball bearings.
For mounting of the ball bearings it is not required to secure them via Loctite again.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
B
A
Figure 97
Figure 98
Figure 99
Chapter 3.5 / 41
Agfa Company Confidential
A