AGFA CR 30-X Service Manual page 265

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(1)
Switch off the digitizer.
(2)
Disconnect the mains cable.
(3)
Simultaneously press the two
buttons (A) positioned underneath
the feed table.
(4)
Pull out cassette unit (B).
(5)
Remove the front cover.
MECHANICAL ADJUSTMENT:
(1)
Insert a cassette the right way, i.e.
black tube side to the top and the
shutter inside the digitizer.
(2)
Push the cassette to the right: The
cassette edge sensor has to
switch.
(3)
If the sensor does not switch, open
the fixation screws at the cassette
edge sensor and push it more to
the left. Fasten the fixation screws
and repeat step (2).
(4)
If the sensor switches, continue
with next step.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
A click must be audible.
When using a multimeter, the
resistance has to be close
to 0 Ω.
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Adjustments and Calibrations
B
Right:
Black tube side on top
CR 30-X
Repair and Service
A
A
Figure 22
Figure 23
Chapter 3.6 / 17