DD+DIS150.06E
(16) Remove the toothed belt.
ASSEMBLY:
(1)
Mount in reverse order.
CALIBRATION:
(1)
Perform a stall calibration for the drawer motor.
Refer to chapter 3.6, Adjustments.
(1)
Insert cassette with unexposed IP in the digitizer.
Verification:
(2)
At the workstation select examination type "System diagnosis – Flatfield".
(3)
Select "ID" to identify the image: Scanning starts.
(4)
Confirm that the image arrives at the workstation.
(5)
Reject the image.
The Toothed Belt Stepper Motor Drawer is replaced.
Result:
2.23

Replacing the Toothed Belt Stepper Motor Roller

REQUIRED SPARE PART*:
CM+9 5175 9050 0
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures

DOCUMENT CONTROL NOTE:

CR 30-X
Type 5175 / 100/110
Repair and Service
517535aj.cdr
Figure 191
Figure 192
Chapter 3.5 / 77
Agfa Company Confidential