DD+DIS150.06E
Solved Image Faults:
In software CRE_1405 it was possible that regular, bright, wide stripes in fast scan
direction appeared, if first a 35x43 cm Genrad scan followed by a 35x43 cm FLFS
(Full leg / full spine) scan a was performed.
8.3

Open Issues

In software CRE_1509 it is erroneously possible to perform the service function
"Diagnose Cycle" and "Endurance Run" remotely. In previous software versions these
two functions were blocked for service connections via "remote desktop".
CAUTION:
A local service engineer may work at the device while a remote service engineer
executes a "Diagnose Cycle": Risk of injury.
Do not execute service function "Diagnose Cycle" via "Remote Desktop" connection.
9

Software Release CRE_1511

Type of new Software*
Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
*Note:
On MedNet, GSO Library always the latest available software release can be
downloaded.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
Software Upgrade
07/2009
5175/100 SN ≥ 5570 and type 5175/110 SN ≥ 6377
MEDNET, GSO Library* (25 MB size)
Software CRE_1511 has only been released with
processing station software NX 2.0.6805 SU1 and SU2.
Together with software upgrade CRE_1511 it is
mandatory to upgrade to NX 2.0.6805 SU1 or SU2, too, if
not yet done earlier.
DD+DIS131.09E, Service Bulletin No. 28, Laser Beam
Detection Toolkit CM+9 5175 2940 0 and FireWire
Connector Shield CM+9 5175 1081 0 available
(Intranet Link
/
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
Extranet
Link)
Agfa Company Confidential