Replacing The Chip Reader - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(13) Open cable holders.
(14) Open 3 screws at the lower right hand side.
(15) Remove the cassette unlocking unit.
ASSEMBLY:
(1)
Mount in reverse order.
CALIBRATION:
(1)
Perform a stall calibration for the locking motor.
Refer to chapter 3.6, Adjustments.
(1)
Insert cassette with unexposed IP in the digitizer.
Verification:
(2)
At the workstation select examination type "System diagnosis – Flatfield".
(3)
Select "ID" to identify the image: Scanning starts.
(4)
Confirm that the image arrives at the workstation.
(5)
Reject the image.
The cassette unlocking unit is replaced.
Result:
2.7

Replacing the Chip Reader

REQUIRED SPARE PART*:
CM+9 0486 1417 0
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 82
Figure 83
Chapter 3.5 / 34
Agfa Company Confidential