AGFA CR 30-X Service Manual page 269

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
NOTE:
Short description of the laser beam position check:
The laser beam is switched on via service client.
By putting the switch on the distributor board to ON position the laser beam
position check functionality is activated: The yellow "LED 1" lights up.
If "LED 2" and "LED 3" on the distribution board are light up, the laser beam
position is OK.
Short description of the laser beam position adjustment:
The laser beam is switched on via service client.
By putting the switch on the distributor board to ON position the laser beam
position check functionality is activated: The yellow "LED 1" lights up.
By opening the screws at the two Optic Module Position Plates the Optic Module
can be shifted this way that both Laser Beam Position Detection Sensors are
covered by the laser beam. In this case LED 2 and LED 3 are light up.
LED 3
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Optic Module
Position Plate 2
EOL Sensor
Laser Beam Position
Detection Sensor 1
LED 2
Distributor Board
Version 2
LED 1
Switch
Figure 26: Functional principle of the laser beam position check and adjustment
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Adjustments and Calibrations
Laser Beam Position
Detection Sensor 2
CR 30-X
Repair and Service
Optic Module
Optic Module
Position Plate 1
BOL Sensor
Calibration Board
517502by.cdr
Version 2
Chapter 3.6 / 21
Agfa Company Confidential