AGFA CR 30-X Service Manual page 429

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS172.06E
The leakage current limits for which this device is tested, excludes installation in the
patient vicinity. Therefore the equipment must be placed outside the patient vicinity
according to the local valid regulation (UL 60601-1 or EN 60601-1).
The patient vicnity is defined as follows:
UL 60601-1:
o 1,83 m (6 feet) beyond the perimeter of the bed (examination table, dental chair,
treatment booth, and the like)
o vertical 2,29 m (7.5 feet) above the floor
EN 60601-1:
o 1,5 m (4.9 feet) beyond the perimeter of the bed (examination table, dental
chair, treatment booth, and the like)
o vertical 2,5 m (8.2 feet) above the floor
Other equipment that is in contact with the patient is not connected directly
to the Agfa system without additional protective measures.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2010
R
Patient vicinity
R = 1.5 m / 4.9 feet (EN 60601-1) or
h = 2.5 m / 8.2 feet (EN 60601-1) or 2.29 m / 7.5 feet (UL 60601-1)
Figure 1: Patient vicinity according to IEC 60601-1 / UL 60601-1
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Installation Planning
Patient vicinity
h
patient_vicinity.cdr
1.83 m / 6 feet (UL 60601-1)
Agfa Company Confidential
1
2
3
Chapter 11 / 27