DD+DIS150.06E
2

LEDS, Fuses and Switches on the Scan Control Board

LED
Red LD2
Switch
S1 / 1
S1 / 2
S2
IMPORTANT:
Fuses are displayed only to check fuses in case of a suspected defective fuse.
Always replace the board if a fuse is defective.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
Electrical and mechanical Codes, Fuses, LEDs
T1A
F1
Figure 2
Meaning
ON if power for the board available.
Meaning
ON = normal operation (= default)
OFF = debug position. Only for R&D purpose.
Not used.
Board reset. Only for R&D purpose.
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
ON
S1
1 2
ON
LD2
T1A
F1#1
Chapter 3.4 / 7
Agfa Company Confidential