AGFA CR 30-X Service Manual page 104

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(3)
Exit the NX software via "NX Main Menu  Exit NX".
(4)
Exit the "Error Viewer".
(5)
Switch off the digitizer.
Deinstall the digitizer software on the control PC: Select "Start  Settings 
(6)
Control Panel  Add or Remove Programs  CR 30-X".
(7)
If at the end of the deinstallation process of the old software a window pops up,
telling: "You must restart your system ...." continue with section 2.1.1 next page.
Otherwise continue with next step.
(8)
Copy file "Setup CR 30-X_xxxx.msi" (xxxx = software version) from USB stick or
CD ROM to drive D: on the control PC.
(9)
In the windows explorer double-click on file Setup CR 30-X_xxxx.msi.
(10) Select <everyone>. See Figure 4.
(11) Click <next> and follow the instructions of the software installer.
(12) When finished, restart the control PC.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Figure 4
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Figure 3
Agfa Company Confidential
Chapter 3.2 / 8