AGFA CR 30-X Service Manual page 378

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS171.06E
IMPORTANT:
After cleaning, wait for about 10 minutes to allow the solvents to evaporate.
(7)
Put the tray with the image plate
back into the cassette.
(8)
Verify that the white phosphor side is
oriented to the black tube side of the
cassette and that the shutter does
not scratch the image plate.
(9)
Mount the Cu filter at the modality.
(10) Place the cassette in parallel to the
rotating anode.
(11) Expose the cassette with 10 µGy.
NOTE:
Select the following exposure parameters using a 1.5 mm Cu filter to obtain
a dose of 10 µGy:
12 mAs
75 kVp
1.3 m (51,2") distance and
large focus.
The values are approximate and may vary among the X-ray devices.
(12) Turn the cassette by 180°.
(13) Again expose with 10 µGy.
(14) Remove the CU filter from the modality.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
Rotating
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
White phosphor side
1
2
Black tube side of the cassette
517509af.cdr
Figure 6
X-Ray tube
Anode
Cathode
517509aa.cdr
OK
Figure 7
Agfa Company Confidential
NOT OK
Chapter 9 / 12