DD+DIS150.06E
(12) Remove 3 screws of stepper motor roller.
(13) Remove motor.
ASSEMBLY:
(1)
Mount in reverse order. Observe IMPORTANT note below.
(1)
Switch on the digitizer.
Verification
(2)
Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3)
Insert cassette with unexposed IP in the digitizer.
(4)
At the workstation select examination type "System diagnosis – Flatfield".
(5)
Select "ID" to identify the image: Scanning starts.
(6)
Confirm that the image arrives at the workstation.
(7)
Reject the image.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
IMPORTANT:
Mount the motor with the motor cable
outlet as shown in the figure enclosed:
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Replacements / Repair Procedures
CR 30-X
Repair and Service
Figure 139
Figure 140
Chapter 3.5 / 58
Agfa Company Confidential