Replacing The Input Rollers - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
ASSEMBLY:
(1)
Mount in reverse order.
NOTE:
To support that the ball bearing fits in its
rail, lift the inner side plate slightly up.
(1)
Switch on the digitizer.
Verification:
(2)
Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3)
Insert cassette with unexposed IP in the digitizer.
(4)
At the workstation select examination type "System diagnosis – Flatfield".
(5)
Select "ID" to identify the image: Scanning starts.
(6)
Confirm that the image arrives at the workstation.
(7)
Reject the image.
The transport unit ball bearing is replaced.
Result:
2.11

Replacing the Input Rollers

The replacement instructions for the Input Rollers, spare part number*
CM+9 5175 9030 2, are enclosed to the spare part.
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the actual revision of the spare part is delivered.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.