DD+DIS150.06E
(10) Log-off the current user at the control PC.
(11) Login as user "crservice".
(12) Select menu:
<Start - Programs - Agfa CR 30-X - Service Client>
(13) Enter user name and password.
(14) Select menu:
<Analysis and Repair – Check laser beam position>:
A window "ESF Initialization" opens and stays for approximately one minute.
See Figure 30.
WARNING:
Laser light. Eye injuries possible.
Do not keep any tool in the laser beam while the laser is switched on.
(15) Select the button <On> to switch on the laser beam. See Figure 31.
The status message shows <Polygon is spinning. Laser ready>.
See Figure 32. The Polygon motor is audible.