Installing The Service Client On The Control Pc - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
NOTE:
It is possible to use the CR30-X service client also on the service PC.
Following functions can be accessed:
Directly after installation of the Service Client:
Function "Explain Error Code"
After Restore of a CR30-X backup file:
View Infocounter
View E-Label entries
View Service Activity Log
View Error Statistics
Prerequisites for Installation:
Windows XP (home or professional)
.Net Visual J# (to be obtained via www.microsoft.com)
2.1

Installing the Service Client on the Control PC

NOTE:
All steps for software upgrade / update of the digitizer are described in a document
which is part of the software on MEDNET GSO Library (readme or enclosure
document). Depending on the software version, the procedure might be different to
the steps described here.
The software installation for the digitizer is made in 2 steps:
Step 1: De-install the previous digitizer software (if applicable) and install the
Step 2: Upload the software from control PC to the digitizer.
IMPORTANT:
Before starting any software de-installation activity at the control PC, switch off the
digitizer. Otherwise the digitizer driver on the control PC will automatically
be de-installed.
(1)
Select "START - Log Off <current user>" to log-off the current user.
(2)
When the log-in screen appears login as "crservice".
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
current software (incl. service client = service software) on the control PC.
Details see this section.
Details see section 2.2.7.
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Tools and auxiliary Means
CR 30-X
Repair and Service
Chapter 3.2 / 7
Agfa Company Confidential