2.2.13

Service Client Menu "Stall Calibration"

The "Stall Calibration" menu allows performing a stall calibration for a specific motor
Purpose
or for all motors.
(1) Select "Preventive Maintenance – Stall Calibration".
Usage
(2) Select the required motor or all motors.
(3) Wait till the stall calibration is finished: Takes approximately 1 minute per motor.
For details see also chapter 3.6, Adjustments and Calibrations.
NOTE:
For troubleshooting of a module the final message parameter "iRunCal" can be used
(see Figure 20):
If executed three consecutive times, the displayed "iRunCal" value may vary ± 2
digits only (e.g. iRunCal values 56, 58, 60). Larger deviations show mechanical
problems in the mechanics of the respective drive.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Figure 20
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 25