DD+DIS150.06E
2.2.16.1
Replacement Scenarios
For exchange of spare parts with E-Labels* two scenarios can be distinguished:
*E-Labels contain spare part specific data, entered in production. Following boards have E-Labels
included: PMT module, Printed circuit board in Optic Module, IP handling board, Slow scan board, PMI /
PMI2 / PMI 3 board.
Item
Type of spare part with
E-Label
Serial number of spare
part written to E-label
Required E-label read/
write direction after part
replacement
Required selection in
menu "Configuration –
Exchanged
Components"
* If a board was used in a digitizer for test purposes and has been put back in the supply chain it will
have a serial number different from 9999.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Scenario 1
Spare part is a sub-
module of a larger
(adjusted) module and
thus not adjusted in
production:
Slow scan board
IP handling board
PMI / PMI2 / PMI 3
Board
9999*
Module specific data
have to be written from
control PC to the board.
Board
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Scenario 2
Spare part is an adjusted
module:
Optic Module
Slow scan unit (incl. Slow
scan board)
PMT with light collector
Cassette Fixation Unit
(incl. IP handling board)
Spare part specific SN,
e.g. 2345
Module specific data have to
be read and copied from new
board to control PC.
Module
Agfa Company Confidential