1

Software Release CRE_1010A

Type of new Software
Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
1.1

Supported Functions

All functions as specified:
Open the inserted cassette (15 x 30 cm; 18 x 24 cm; 24 x 30 cm; 35 x 43 cm) and
transport the IP to the Scan Unit.
Acquire the ID data via fast ID mode from the processing station.
Scan (= stimulate and scan) the image on the IP.
Send the raw image to the processing station.
Erase the IP.
Put the IP back in its cassette, close cassette.
Update ID data on IP (cycle counter).
1.2

Limitations

Shading calibration curves are stored at the processing station only. This means, if a
digitizer is operated with another processing station without restore, its shading
calibration is lost.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
1st customer release
09/2006
 1001
MEDNET, GSO Library* (8 MB size)
None
DD+DIS365.06E, Service Bulletin No. 01, CR 30-X
System released for Sales
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
(Intranet Link
Agfa Company Confidential
/
Extranet
Link)