DD+DIS150.06E
CAUTION:
If switch on distributor board stays in "ON" position during digitizer operation,
reduced image quality possible.
Set the switch on the distributor board to the OFF position before handing the digitizer
over to the customer.
(18) Set the switch on the distributor
board to the OFF position.
(19) Mount all covers.
(20) Connect the cables in the rear.
(21) Continue with section 2.5.3.
The Laser Beam Position Check is performed.
Result
2.5.2
Performing the Laser Beam Position Adjustment
NOTE:
Before starting the laser beam position adjustment, first perform the laser beam
position check. See steps (1) to (17) in section 2.5.1.
(1)
Open the four optic module fixation
screws.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
ON
OFF
Figure 34
Figure 35
Chapter 3.6 / 25