Set Digitizer Operational; Checking The Technical Image Quality Of The System; Checking Image Plate Before Exposure - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
6

Set Digitizer operational

The purpose of this section is to show the tasks that are required to set the digitizer
operational.
The following table lists these tasks.
Task
Check technical image
quality of the digitizer.
*Complete required time per Task Code to be entered in the Agfa repair report.
6.1

Checking the Technical Image Quality of the System

During the "Technical Image Quality Check" of the system a flatfield image is exposed
and checked at the lightbox. The procedure is divided in following steps:
#
Task
1
Clean and check the image plate.
2
Erase the cassette.
3
Expose the cassette.
4
Scan the flatfield.
5
Check the image at the lightbox or viewing station.
6.1.1

Checking Image Plate before Exposure

(1)
Select the largest available cassette. If possible, use a 35x43 cm Genrad
cassette.
If in size 35x43 cm only FLFS cassettes are available, use a FLFS cassette.
Distinction between Genrad and FLFS cassette see Figure 22 on page 21.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
Required
Importance
Time
30 minutes
mandatory
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Task Code*
Section
INS
Section
6.1.1
6.1.2
6.1.3
6.1.4
6.1.5
Chapter 1 / 16
Agfa Company Confidential
6.1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents