Performing Test Cycles
(1)
In the CR30-X service client select:
Analysis & Repair → Endurance
Run Cycles
(2)
Carry out five test cycles with each
available cassette format.
4.4
Technical Image Quality Check of the System
Note:
The "Technical Image Quality Check" of the system is divided in 3 steps:
1. Creating a flatfield image
2. Checking the image quality
3. If the image quality is not OK: Comparing the flatfield with the test patterns "lowest
acceptable quality"
4.4.1
Creating a Flatfield Image
(1)
Take the same cassette you have
used for image quality check at the
beginning of the maintenance
procedure.
(2)
Mount the CU filter at the modality.
(3)
Place the cassette in parallel to the
rotating anode.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
Rotating
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Figure 24
X-Ray tube
Anode
Cathode
517509aa.cdr
OK
Figure 25
Agfa Company Confidential
NOT OK
Chapter 9 / 23