DD+DIS150.06E
(1)
Connect the power cord.
Verification
(2)
Switch on the digitizer.
(3)
Open the cassette unit approximately 1 cm.
(4)
Confirm that the switch opens and switches off the digitizer.
(5)
Close the cassette unit.
(6)
Wait till the green LED lights up.
The cassette unit safety switch is adjusted.
Result
2.4
Cassette Edge Sensor Position Adjustment
Adjustment of the switch position, so that it reliably detects whether the cassette is
Purpose
inserted the right way.
After replacement of the cassette edge sensor
Use Case
REQUIRED TOOLS:
REQUIRED TIME:
20 minutes
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Optional:
1,5 mm
200
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Adjustments and Calibrations
Multimeter
(Ohmmeter)
Agfa Company Confidential