Software Patch Cre_1206C; Solved Bugs With Software Cre_1206C Compared To Cre_1206A - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
3

Software Patch CRE_1206C

Type of new Software
Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
*Note:
On MedNet, GSO Library always the latest available software release can be
downloaded.
3.1

Solved Bugs with Software CRE_1206C compared to CRE_1206A

Solved wrong image plate sensitivity which was sent from the digitizer to the NX
workstation resulting in too high displayed lgM-values.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
Software Upgrade
01/2007
 1021
MEDNET, GSO Library* (600 kB size)
Software CRE_1206 has to be installed before installing
the software patch CRE_1206C.
DD+DIS001.07E, Service Bulletin No. 05, Displayed
LgM-Values too high: Install Software Patch CRE_1206C
(Intranet Link
/
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
Extranet
Link)
Agfa Company Confidential
Chapter 3.8 / 7