AGFA CR 30-X Service Manual page 106

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
Digitizer was not switched off during software de-installation on the control PC. In this
Cause
case the digitizer driver on the control PC will also be de-installed.
(1)
Cancel the "Found new Hardware Wizard" and continue with software
Solution
installation.
(2)
Install the driver via "found new hardware wizard" which pops up automatically
after software upload to the digitizer (and restart of the control PC). See Figure 7.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Figure 7
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 10