AGFA CR 30-X Service Manual page 55

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
IMPORTANT:
If you use a FLFS cassette for flatfield exposure, the image will be slightly darker in
the upper and lower image area where the backscatter protection is removed (approx.
1 cm each). These darker zones also have to be used for image quality evaluation in
slow scan and fast scan direction.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
Flatfield with Genrad Cassette
Genrad Cassette
Backscatter
Protection
Figure 23: Main differences of Genrad and FLFS Cassettes
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Flatfield with FLFS Cassette
Effect exaggerated
Figure 22
FLFS Cassette
FLFS
Agfa Company Confidential
FLFS
10 mm
10 mm
Chapter 1 / 21