NOTE:
All modules with E-labels have installation instructions enclosed, which describe the
correct proceeding.
8.2

Solved Bugs with Software CRE_1509 compared to CRE_1405

Software CRE_1509 contains following bug-fixes:
8.2.1

Misleading Error Messages in the Logfiles (HQ_0702280001)

Symptom
Digitizer logfiles contains misleading, obsolete error messages
Defect
Software problem
Solution
Problem fixed in software CRE_1509
8.2.2
Missing automatic E-Label data Synchronization after Replacements
(HQ_0704270003)
Symptom 1
After replacement of an IP Handling Board (CM+9 5175 5230 X) it is possible that after
insertion of the cassette, during cassette clamping error 4152 pops up "Cassette fixer
detected unknown format".
Symptom 2
After replacement of a Slow Scan Board (CM+9 5175 2120 X) it is possible that the
aspect ratio of the scanned image is not exactly 1:1 anymore, but e.g. 1:0,99 or 1:1,01.
IMPORTANT:
Both symptoms only appear, if both of following conditions are fulfilled:
The digitizer at the production site which is used for board testing and the digitizer
on site, where the new boards are built in, deviate a lot in mechanical tolerances
concerning cassette clamping or the slow scan gear box.
The NX processing station is always switched on.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
Agfa Company Confidential