AGFA CR 30-X Service Manual page 279

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(16) Wait until the calibration data is transferred. In case of an error refer to section
3.1.2, Possible Reasons for a failed Shading Calibration.
(17) Remove cassette from digitizer.
NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
The digitizer records 6 shading calibration lines (Gain 1 to Gain 6), one for
each facet of the polygon mirror.
Additionally to the storage on the PMT, the calibration file is stored in the
following folder:
C:\Documents and Settings\All Users\Application
Data\Agfa\CR30-X\Gains00.stc
The calibration file is overwritten by the new one. To keep a certain
calibration file, save it with a different name.
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
Adjustments and Calibrations
Figure 43: Example screenshot
CR 30-X
Repair and Service
Chapter 3.6 / 31
Agfa Company Confidential