DD+DIS150.06E
CAUTION:
The nuts of the cassette edge sensor mounting can be lost very easily.
Only loosen the screws of the cassette edge sensor approx. 1 to 2 turns.
Always have a "small parts kit*" CM+9 5175 9070 1 with you: This contains 2 spare
screws and nuts for the cassette edge sensor mounting.
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the actual revision of the spare part is delivered.
(12) Remove cables (1) at cassette edge
sensor.
(13) Open the two screws (2) approx. 1 to 2
turns.
(14) Slide out the cassette edge sensor (3).
ASSEMBLY:
(1)
Mount in reverse order. See important note below.
IMPORTANT:
Pin 2 is not connected at the cassette edge
sensor.
MECHANICAL ADJUSTMENT:
(1)
Perform a "cassette edge sensor" adjustment. Refer to chapter 3.6, adjustments.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
2
1
Figure 49
1
3
2 (not connected)
Chapter 3.5 / 22
Agfa Company Confidential
1
3