DD+DIS150.06E
2.2.6
Service Client Menu "HOT Report" (Hand-over Test Report)
The menu "Hot Report" gives a summary of the last scan cycle. The main steps of this
Purpose
cassette and IP cycle are listed.
It can be used to check e.g. during installation, if the performed scan cycle was
successfully.
The HOT report is overwritten with each new scan cycle.
(1) Select "Reporting - HOT Report".
Usage
(2) Select "Save as" to save it on a backup medium (e.g. hard disk or USB stick).
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Figure 13: Example for a HOT report
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 17