DD+DIS150.06E
(3)
Mount the PMT gauge on the light
collector.
(4)
Check the distance between PMT
gauge and PMI carrier (see two
circles in Figure 17):
The PMI carrier should just touch
the PMT gauge.
(5)
If the distance is OK, continue
with step (9), otherwise continue
with next step.
(6)
Open the four adjustment screws.
See arrows in Figure 17.
(7)
Adjust horizontal position as well
as the angle of the PMI board
carrier.
(8)
Fasten the four adjustment
screws.
(9)
Remove the PMI carrier.
(10) Remove PMT gauge.
(11) Mount PMT on the light collector.
(12) Keeping the PMT in its mounting
position by hand, mount the PMI
board carrier. See Figure 18.
(13) Re-assemble the digitizer.
(14) Perform a shading calibration.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 16
Figure 17
Figure 18
Chapter 3.6 / 13