AGFA CR 30-X Service Manual

AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

H
C
EALTH
ARE

Service Manual

Imaging Services
Document No: DD+DIS150.06E
CR 30-X
Type 5175/100/105/110/120/125
st
1
Edition *
*
For the revision status per chapter please refer to the latest version of the order list in the GSO Library.
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
11-2010 printed in Germany
Document Node ID: 14301131
eq_00_about manual_e_template_v07
Agfa Company Confidential
Copyright © 2010 Agfa HealthCare N.V.
1

Advertisement

Table of Contents
loading

Summary of Contents for AGFA CR 30-X

  • Page 1: Service Manual

    The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 11-2010 printed in Germany Document Node ID: 14301131 eq_00_about manual_e_template_v07 Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
  • Page 2 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 3 Added digitizer subtypes 105, 120, and 125. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 0 / 4 11-2010 Type 5175/100/105/110/120/125...
  • Page 4 Service Bulletins Installation Planning Glossary DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 0 / 5 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 5 In this example a file name has to be entered as term. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 0 / 6 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 6: About This Manual

    Digitizer (ex factory no name on the covers) All of the above mentioned subtypes are called "digitizer" in this document. The functional differences between the different subtypes are described in the CR 30-X Installation Planning Document, DD+DIS172.06E. The digitizer is part of a system comprising following components: ...
  • Page 7 This Generic Safety Directions document comprises the general safety relevant information including relevant environmental and occupational safety instructions for the Service Engineer. It is valid for all Agfa HealthCare Imaging Products and part of each Service Documentation as well as Installation Planning document.
  • Page 8 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 9: Table Of Contents

    SAFETY DIRECTIONS FOR REMOTE SERVICE ACTIVITIES ......26 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 3 of 07-2009 Agfa Company Confidential...
  • Page 10 ERASING PROTECTED HEALTH INFORMATION (PHI) ........28 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 4 of 07-2009 Agfa Company Confidential...
  • Page 11: Disclaimer

    Fitters, engineers and other persons who are not employed by or otherwise directly affiliated with or authorized by Agfa HealthCare or one of its affiliates are directed to contact one of the local offices of Agfa HealthCare or one of its affiliates before attempting installation or service procedures.
  • Page 12: Used Icons

    Is additional information without influence on the action or step! DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 6 of 07-2009 Agfa Company Confidential...
  • Page 13: Labels

    C&W _005.cdr DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 7 of 07-2009 Agfa Company Confidential...
  • Page 14: System Labels Concerning Laser Radiation

    CLASS 2 LASER PRODUCT conditions. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 8 of 07-2009 Agfa Company Confidential...
  • Page 15 CLASS 4 LASER PRODUCT DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 9 of 07-2009 Agfa Company Confidential...
  • Page 16: Product Complaints

    Agfa HealthCare must be notified immediately by telephone, fax or written correspondence to the following address: Agfa Service Support - local support addresses and phone numbers listed on: www.agfa.com Agfa – Gevaert N.V.
  • Page 17: Intended Use

    DD+DIS238.06E Generic Safety Directions Intended Use This Agfa HealthCare product should only be operated in a hospital or clinical radiological environment by qualified staff. It must only be operated according to its specifications and its intended use. Any operation not corresponding to the specifications or intended use may result in hazards, which in turn may lead to serious injuries or fatal accidents (for example electric shocks).
  • Page 18: Environmental And Occupational Safety Instructions

    Is obligated to adhere strictly to regulations and instructions. • Shall familiarise himself or herself with the provisions of the Agfa Healthcare Health, Safety and Environment Policy and any specific rules or procedures relating to occupational safety at work and the protection of the environment.
  • Page 19: Connections To Other Equipment

    Changes or additions to the equipment must only be carried out by persons authorized to do so by Agfa HealthCare. Such changes must comply with best engineering practice and all applicable laws and regulations that have the force of law within the jurisdiction of the hospital.
  • Page 20: Accessories And Spare Parts

    • Use only tools and measuring instruments which are suitable for the procedure. • Only approved Agfa HealthCare accessories must be used. For a list of compatible accessories contact your local Agfa HealthCare organization or www.agfa.com. Compliance Directive for HealthCare Imaging Products:...
  • Page 21 It is a multilateral agreement among participating countries and certification organizations. Agfa has produced a CB test report and claims national certification in all other member countries of the CB Scheme.
  • Page 22 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 16 of 07-2009 Agfa Company Confidential...
  • Page 23: Safety Directions For Operation

    Accessibility of the mains power switch: CAUTION: Do not obstruct the mains power switch. Position the Agfa HealthCare product so that it is possible to disconnect the mains power connection. • Under certain conditions the Agfa HealthCare product will show a display containing a message.
  • Page 24: Radiation Protection

    Disconnect the power supply from the equipment prior to cleaning the equipment. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 18 of 07-2009 Agfa Company Confidential...
  • Page 25: General Safety Directions For Service Activities

    If "Loctite" has to be used to secure screws, this is stated in the text. • Any Agfa service PC or tool which is to be connected via RS232, RJ45, USB or other interface to an Agfa device must not be connected to the mains but must be operated on its internal battery or indirect supply (low voltage).
  • Page 26: Safety Directions For Installation Planning Activities

    Electrical device. Shock possible. INSTRUCTION: • Do not position Agfa EQUIPMENT so it is difficult to operate the disconnection device when an APPLIANCE COUPLER or separable plug is used as isolation. • Local and International wiring regulations must be observed. Check all supplies and voltages, currents, trips and fuses with the Hospital facilities department or their engineers.
  • Page 27 Hospital management is responsible for the position, location and fixing of all equipment. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 21 of 07-2009 Agfa Company Confidential...
  • Page 28: Safety Directions For Installation Activities

    If not otherwise stated, installation and configuration is performed by Agfa HealthCare trained personnel. • If damage of the package is visible from the outside contact your local AGFA representative. • Apart from wearing the required protective clothing, e.g. safety boots and gloves, care must be taken that heavy loads are correctly lifted/carried to avoid injury.
  • Page 29: Safety Directions For Maintenance And Repair Activities

    Safety Directions for Maintenance and Repair Activities • This Technical Documentation identifies the parts on which preventive inspection and maintenance shall be performed by Agfa HealthCare service personnel, including the periods to be applied. • In general the device has to be switched off during service activities. Exception: If the device is switched on to perform tests pay particular attention to any hazards due to moving and rotating parts.
  • Page 30 NOTE: On MedNet, GSO Library path "General Info => Agfa HealthCare => Publications => Service Manual" a form “IEC 62353:2007 Test Documentation” for CR Digitizers is available, to be handed out to the responsible organization...
  • Page 31 When cleaning optical elements follow the service manual precisely. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 25 of 07-2009 Agfa Company Confidential...
  • Page 32: Safety Directions For Remote Service Activities

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 26 of 07-2009 Agfa Company Confidential...
  • Page 33: Generic Safety Directions For Healthcare Imaging Products

    The Recycling Passports are meant to be used as information for waste treatment partners and companies that want to recycle end-of-life Agfa equipment. To get a copy of the required Agfa HealthCare Recycling Passport please contact your local Sales organization.
  • Page 34: Waste Disposal

    For more detailed information about take-back and recycling of this product, please contact your local Agfa service organization. By ensuring this product is disposed of correctly, you will help prevent potential negative consequences for the environment and human health, which could otherwise be caused by inappropriate waste handling of this product.
  • Page 35 ► Referenced Documents Document Title DD+DIS287.09E CR 30-X Service Bulletin No. 30 - Software CRE_1607 available DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5...
  • Page 36 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 37 INSTALLATION CHECKLIST ....................24 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 3 11-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 38: Scope Of This Document

    Prepare required tools. 5 minutes mandatory *Complete required time per Task Code to be entered in the Agfa repair report. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
  • Page 39: Checking The Installation Site Prerequisites

    Checking the Installation Site Prerequisites Confirm that all prerequisites described in the "Installation readiness checklist" are fulfilled. This checklist is available at the end of CR 30-X Service Documentation Chapter 11, Installation Planning. Preparing the required Tools Following tools are required during the installation procedure:...
  • Page 40: Inspection And Unpacking

    Check shipment 5 minutes mandatory completeness. *Complete required time per Task Code to be entered in the Agfa repair report. Checking the Packing for obvious Damage (1) Check the packing material for visible transport damage such as:  Dented edges ...
  • Page 41: Unpacking The Digitizer

    Lift the digitizer from the pallet on the provided table. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 7 11-2009 Type 5175 / 100/110...
  • Page 42: Checking The Shipment Completeness

    2 minutes mandatory Switch on the digitizer. 3 minutes mandatory *Complete required time per Task Code to be entered in the Agfa repair report. Mounting the Distance Holder REQUIRED PARTS (part of delivery): Figure 2 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
  • Page 43: Connecting The Cables

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 9 11-2009 Type 5175 / 100/110...
  • Page 44 Man ufacturer AGFA H EALTH CAR E N . V . Septe straat 27 - 26 40 - Morts el - BE L G IUM M ED I C AL ELEC TRI CA L...
  • Page 45: Switching The Digitizer On

    Table 1: Description of Status Indicator DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 11 11-2009 Type 5175 / 100/110...
  • Page 46: Configuring Digitizer Settings

    5 minutes optional Indicator. *Complete required time per Task Code to be entered in the Agfa repair report. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5...
  • Page 47: Filling In Of Site Specific Data

    DD+DIS150.06E Controls, Connections, and Setup Procedures Filling in of Site specific Data On the control PC select: <Start - Programs - Agfa CR 30-X - Service Client> Enter user name and password. Select: <Configuration - Site Specific Data> Enter the site specific data.
  • Page 48: Checking Software Version

    PC. (1) Log-off the current user at the control PC. (2) Login as user crservice. (3) Select menu: <Start - Programs - Agfa CR 30-X - Service Client> (4) Enter user name and password. (5) Select <Get installed version>.
  • Page 49: Configuring Maintenance Indicator

    DD+DIS150.06E Controls, Connections, and Setup Procedures Configuring Maintenance Indicator In the CR 30-X Service Client select <Maintenance Indicator>. Configure whether maintenance indicator should be activated shown or not (default = off).  "On" (checkbox selected) means: Customer will be reminded daily or when starting the Control PC to perform about a preventive maintenance via following message: "Maintenance interval will expire within a short time.
  • Page 50: Set Digitizer Operational

    *Complete required time per Task Code to be entered in the Agfa repair report. Checking the Technical Image Quality of the System During the “Technical Image Quality Check” of the system a flatfield image is exposed and checked at the lightbox.
  • Page 51 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 17 11-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 52: Erasing The Cassette

    Remove the cassette with the erased image plate from the cassette output. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 18 11-2009 Type 5175 / 100/110...
  • Page 53: Creating The Flatfield Test Image

    Figure 21 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 19 11-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 54: Scanning The Flatfield Image

    For external partners accessing the “Troubleshooting Guide” on MedNet, GSO Library may be restricted. In such a case please contact your local Agfa representative. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
  • Page 55 Figure 23: Main differences of Genrad and FLFS Cassettes DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 21 11-2009 Type 5175 / 100/110...
  • Page 56: Completing The Digitizer Installation

    Clinical Application Specialist. *Complete required time per Task Code to be entered in the Agfa repair report. Performing a Backup (1) In the CR 30-X service client select the menu <Backup & Restore>. (2) Keep the default settings as shown in Figure 24.
  • Page 57: Hand-Over Of The System To The Clinical Application Specialist

    Any irregularities during the installation. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 23 11-2009 Type 5175 / 100/110...
  • Page 58: Installation Checklist

    14. Hand-over of the System to the Clinical Application Specialist DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 1 / 24 11-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 59 DD+DIS150.06E Functional Description Chapter 2 List of Contents CR30-X Digitizer as a part of the CR 30-X System ....1 System Overview .................... 1 CR30-X in a Nutshell..................1 NX Workstation in a Nutshell................. 2 Image Plates with IPs in a Nutshell ............... 3 Workflow of the CR 30-X System..........4...
  • Page 61: Cr30-X Digitizer As A Part Of The Cr 30-X System

    DD+DIS150.06E Functional Description CR30-X Digitizer as a part of the CR 30-X System System Overview The CR30-X digitizer is part of the Digitizer NX Workstation "CR 30-X System" which comprises: Cassette • Digitizer CR30-X • NX Workstation • Cassette with Image Plate...
  • Page 62: Nx Workstation In A Nutshell

    The CR30-X cannot be run together with ADC QS workstations. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 2 / 2 CR30-X Edition 1, Revision 0 Agfa Company Confidential Type 5175/100...
  • Page 63: Image Plates With Ips In A Nutshell

    Figure 3: CR30-X Cassette with Image Plate "MD40" DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 2 / 3 Type 5175/100 Agfa Company Confidential...
  • Page 64: Workflow Of The Cr 30-X System

    Functional Description DD+DIS150.06E Workflow of the CR 30-X System Take exposure Take cassette for next exposure Insert exposed Cassette in Digitizer Go to processing station and select Patient: The IP* cycle starts. FireWire cable Wait till IP* cycle is finished (approx. 1 min.) 517502ah.cdr...
  • Page 65: Hardware And Software

    DD+DIS150.06E Functional Description Hardware and Software Agfa NEXUS Operating System Agfa NEXUS Operating System Erasure Chip User NXApplication Board Reader Interface Software µ Contr. TWAIN Interface µ Contr. Handling FPGA Board FireWire CR30-X Server FireWire Error Interface Application Interface Viewer...
  • Page 66: Digitizer Boot-Up

    It checks e.g. whether motors and sensors are connected. Figure 6 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 2 / 6 CR30-X Edition 1, Revision 0 Agfa Company Confidential Type 5175/100...
  • Page 67: Scanning

    Calibration Board BOL Sensor 517502ae.cdr Figure 7 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 2 / 7 Type 5175/100 Agfa Company Confidential...
  • Page 68: Description Of Most Important Components For Scanning

    The length of the AD window and "Laser On" depends on the image plate size. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 2 / 8 CR30-X Edition 1, Revision 0 Agfa Company Confidential Type 5175/100...
  • Page 69: Signal Corrections Before And During Scan

    Parameters of REF LUM: Variations of LEDs; LED mounting position. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 2 / 9 Type 5175/100 Agfa Company Confidential...
  • Page 70 The digital value of each pixel is corrected with a correction factor determined during shading calibration. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 2 / 10 CR30-X Edition 1, Revision 0 Agfa Company Confidential Type 5175/100...
  • Page 71 *Location of Connector, E-Label, EEPROM and Flash ROM simplified 517502ad.cdr Figure 9 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 2 / 11 Type 5175/100 Agfa Company Confidential...
  • Page 72: Stall Detection

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 2 / 12 CR30-X Edition 1, Revision 0 Agfa Company Confidential Type 5175/100...
  • Page 73 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 2 / 13 Type 5175/100 Agfa Company Confidential...
  • Page 75 CR 30-X Type 5175/100/105/110/120/125 ► Purpose of this Document  The Software / Hardware Compatibility Chart CR 30-X System informs the Field Service Engineer about released CR 30-X System versions and their components.  This document will be updated regularly.
  • Page 76 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 77 NOTE:  The CR 30-X System version contains a continuous number which is increased when one of the delivered system components changes.  Explanation of symbols in tables next pages:...
  • Page 78 NX 8000 NX 2008 on FLFS phase 2 WHITE NX2008 version Discovery HP DC7800 LABEL DIGITIZER CR 30-X Tpye 5175/100 Digitizer Type 5175/110 Digitizer Model Files Speedclass free, CR30SCF.XML V2.1.0.0 Speedclass operation, CR30.XML V2.1.0.0 Speedclass free, CR30SCF.XML V2.2.0.0 Speedclass operation, CR30.XML V2.2.0.0...
  • Page 79 NX Musica2 Veterinary Small Animals NX Office Viewer System Options Cart for CR 30-X DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X...
  • Page 80 CRE_1405 CRE_1405 CRE_1405 CRE_1509 CRE_1509 CRE_1509 CRE_1509 CRE_1509 ALL-in-1 CART CR 30-X Mobile Kit for CR 30-X CR Full Body Cassette Holder Optional CR Anti-scatter grid version 2 Optional CR EasyLift Cassettes 35x43 CR MD4.0T Genrad + FLFS 35x43 CR MD4.0T General Set 35x35 CR MD4.0T...
  • Page 81 NX 8200 CRE_1703 layer redesign CRE_1607A (not NX 8400 CRE_1901 version introduced) DIGITIZER CR 30-X Tpye 5175/100 CR 30-X Tpye 5175/105 Digitizer Type 5175/110 Digitizer Type 5175/120/125 Digitizer Model Files Speedclass free, CR30SCF.XML V2.1.0.0 Speedclass operation, CR30.XML V2.1.0.0 Speedclass free, ...
  • Page 82 Small Animals NX Office Viewer DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 2 / 8 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 83 = only available for digitizer type 5175/110/120/125 = only available for digitizer type 5175/100/105 = CR 30-X model file V2.2.0.0 is part of NX software for NX ≥ NX 8200 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
  • Page 84 (Intranet link Extranet link) Upgrade possibilities for CR 30-X Digitizer Software:  CR 30-X – Service Bulletin No. 44, DD+DIS308.10E, Software CRE_1901 and new Digitizer Subtypes available (Intranet Link Extranet Link) Upgrade to FLFS application for already installed CR 30-X Systems: ...
  • Page 85 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 04-2009 printed in Germany Document Node ID: 12144667 eq_03-1_safety-repair_e_template_v07 Agfa Company Confidential Copyright © 2009 Agfa HealthCare N.V.
  • Page 86 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 87 Electrical Check after Power Supply Replacements...............11 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 3 04-2009 Type 5175 /100/110...
  • Page 88: Intended Use

    517631ap.cdr Figure 2 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 4 04-2009 Type 5175 /100/110 Agfa Company Confidential...
  • Page 89 U L60601- 1 / Ma d e in Ge rma ny Peiss en berg MMM MM -YYYY 9F 03 FireWire Manufacturer AGFA HEALTHCARE N.V. Septestraat 27 - 2640 - Mortsel - BELGIUM 517531ae.cdr MEDICAL ELECTRICAL Connector EQUIPMENT WITH RESPECT TO...
  • Page 90 Figure 9 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 6 04-2009 Type 5175 /100/110 Agfa Company Confidential...
  • Page 91 CLASS IIIb LASER PRODUCT Figure 10 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 7 04-2009 Type 5175 /100/110 Agfa Company Confidential...
  • Page 92: Safety Notes For Service Activities

    Planning, section "Accessories". DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 8 04-2009 Type 5175 /100/110 Agfa Company Confidential...
  • Page 93: Safety Notes For Installation

    5 1 7 5 3 1 a b . c d r Figure 11 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 9 04-2009 Type 5175 /100/110 Agfa Company Confidential...
  • Page 94: Safety Note For Replacement Of The Power Supply

    Handle the light collector with great care. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 10 04-2009 Type 5175 /100/110...
  • Page 95 Refer to Figure 12. Figure 12 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 11 04-2009 Type 5175 /100/110 Agfa Company Confidential...
  • Page 96 Example for Germany: BGV A3 and DIN VDE 0751 or subsequent regulations. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.1 / 12 04-2009 Type 5175 /100/110...
  • Page 97 2.2.16 and 2.2.17. ► Referenced Documents Document Title DD+DIS131.09E CR 30-X Service Bulletin number 28: Laser Beam Detection Toolkit CM+9 5175 2940 0 and FireWire Connector Shield CM+9 5175 1081 0 available DD+DIS333.09E CR 30-X Service Bulletin number 32: Software CRE_1607A available DD+DIS150.06E...
  • Page 98 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 99 REMOTE DESKTOP ......................34 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 3 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 100: Machine Specific Tools

    When ordering, the actual revision of the spare part is delivered. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 4 03-2010 Type 5175 / 100/110...
  • Page 101: Recommended Tools

    Figure 1 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 5 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 102: Digitizer Service Client

    The up-to-date CR30-X service client is available via MEDNET GSO Library path: Computed Radiography – CR Digitizers – CR30-X  It is started via menu: All Programs - Agfa CR 30-X - Service Client Figure 2: CR30-X Service Client menus DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet.
  • Page 103: Installing The Service Client On The Control Pc

    When the log-in screen appears login as "crservice". DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 7 03-2010 Type 5175 / 100/110...
  • Page 104 Switch off the digitizer. Deinstall the digitizer software on the control PC: Select "Start  Settings  Control Panel  Add or Remove Programs  CR 30-X". If at the end of the deinstallation process of the old software a window pops up, telling: "You must restart your system .."...
  • Page 105: Troubleshooting Software Installation Problems At The Control Pc

    Figure 6 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 9 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 106 Figure 7 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 10 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 107: Service Client Menus

    Backup & Restore DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 11 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 108: Service Client Menu "Info Counter

    Figure 8 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 12 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 109: Service Client Menu "E-Label

    Figure 9 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 13 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 110: Service Client Menu "Sw-Versioning

    Scan Board (SSC Board) have always to be the same. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 14 03-2010 Type 5175 / 100/110...
  • Page 111: Service Client Menu "Service Activity Log

    When the logfile exceeds 1 MB, a new and empty “ServiceActivity.log” file is created. The previous one is automatically renamed to ServiceActivity1.log, which is stored in directory C:\Documents and Settings\All Users\Application Data\Agfa\CR 30-X\Log on the control PC. All “ServiceActivity.log” files are part of a backup.
  • Page 112: Service Client Menu "Error Statistics

    Figure 12 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 16 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 113: Service Client Menu "Hot Report" (Hand-Over Test Report)

    Figure 13: Example for a HOT report DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 17 03-2010 Type 5175 / 100/110...
  • Page 114: Service Client Menu "Sw Upload

    Figure 14 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 18 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 115 Select "Upload selected nodes". Wait for approximately 7 minutes till message "versions are current" appears in the message box. Close the CR 30-X service client. Answer the dialogue popping up “did you perform any repair or maintenance activities” with "Yes".
  • Page 116: Service Client Menu "Diagnose Cycle

    Figure 15 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 20 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 117: Service Client Menu "Explain Error Code

    Figure 16 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 21 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 118: Service Client Menu "Endurance Run Cycle

    Figure 17 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 22 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 119: Service Client Menu "Selftest Of Device

    Figure 18 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 23 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 120: Service Client Menu "Calibration

    Figure 19 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 24 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 121: Service Client Menu "Stall Calibration

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 25 03-2010 Type 5175 / 100/110...
  • Page 122: Service Client Menu "Site Specific Data

    Figure 21 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 26 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 123: Service Client Menu "Device Specific Data

    The device specific data are “read-only”. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 27 03-2010 Type 5175 / 100/110...
  • Page 124: Service Client Menu "Exchanged Components

    Replacement Scenarios and 2.2.16.2, Handling of Replacements in different Software Versions DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 28 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 125 9999. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 29 03-2010 Type 5175 / 100/110...
  • Page 126 2 page). DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 30 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 127: Service Client Menu "Maintenance Indicator

    CRE_1607. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 31 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 128: Service Client Menu "Backup & Restore

    Step 4: Switch on digitizer. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 32 03-2010 Type 5175 / 100/110...
  • Page 129 Figure 25 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.2 / 33 03-2010 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 130: Remote Desktop

    Purpose (1) Connect the service PC to the network where the control PC with digitizer is Usage located (e.g. via Agfa SRSS). (2) On the local PC select ‘Start – Programs – Accessories – Communications – Remote Desktop’. (3) Enter the hostname or IP address of the processing station.
  • Page 131 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 07-2008 printed in Germany Document Node ID: 12152576 eq_03-3_troubleshooting_e_template_v05 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 132 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.3 / 2 07-2008 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 133 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 04-2008 printed in Germany Document Node ID: 12271306 eq_03-4_codings_e_template_v06 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 134 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 135 LEDS ON THE USER INTERFACE BOARD................10 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 3 Agfa Company Confidential 04-2008...
  • Page 136: Leds And Switches On The Pmi / Pmi2 Board

    Green LED, default ON. Voltage +5 V okay. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 4 Agfa Company Confidential 04-2008...
  • Page 137 LED is programmed. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 5 Agfa Company Confidential 04-2008 Type 5175 / 100/110...
  • Page 138 OFF. Only for R&D purpose. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 6 Agfa Company Confidential 04-2008 Type 5175 / 100/110...
  • Page 139: Leds, Fuses And Switches On The Scan Control Board

    Always replace the board if a fuse is defective. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 7 Agfa Company Confidential 04-2008...
  • Page 140: Leds, Fuses And Switches On The Ip Handling Board

    Always replace the board if a fuse is defective. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 8 Agfa Company Confidential 04-2008...
  • Page 141: Fuses On The Pmi Power Distribution Board

    Always replace the board if a fuse is defective. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 9 Agfa Company Confidential 04-2008...
  • Page 142: Leds On The User Interface Board

    • Error DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.4 / 10 Agfa Company Confidential 04-2008 Type 5175 / 100/110...
  • Page 143 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 03-2009 printed in Germany Document Node ID: 12143682 eq_03-5_replacements_e_template_v07 Agfa Company Confidential Copyright © 2009 Agfa HealthCare N.V.
  • Page 144 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 145 Replacing the Toothed Belt Stepper Motor Roller ..............77 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 3 03-2009 Type 5175 / 100/110...
  • Page 146 Replacing the Power Supply....................100 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 4 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 147 Figure 2 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 5 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 148 Figure 6 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 6 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 149: Replacing The Left Cover

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 7 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 150 Figure 11 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 8 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 151: Replacing The Right Cover

    10 min. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 9 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 152 Mount in reverse order. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 10 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 153: Replacing The Top Cover

    Figure 17 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 11 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 154 The top cover is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 12 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 155: Replacements In Cassette Unit

    Figure 20 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 13 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 156 Figure 24 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 14 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 157 Do not bend IP guide plate at removal. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 15 03-2009 Type 5175 / 100/110...
  • Page 158 Figure 31 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 16 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 159 Figure 35 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 17 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 160 Figure 39 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 18 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 161 The Antistatic Brush Cassette Unit Entry is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 19 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 162: Replacing The Cassette Edge Sensor

    Figure 44 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 20 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 163 Figure 48 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 21 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 164 Perform a "cassette edge sensor" adjustment. Refer to chapter 3.6, adjustments. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 22 03-2009 Type 5175 / 100/110...
  • Page 165: Replacing The Cassette Fixation Steel Band

    When ordering, the actual revision of the spare part is delivered. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 23 03-2009 Type 5175 / 100/110...
  • Page 166: Replacing The Cassette Opener Unit

    Figure 51 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 24 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 167 Figure 55 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 25 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 168 Do not bend IP guide plate at removal. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 26 03-2009 Type 5175 / 100/110...
  • Page 169 Figure 62 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 27 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 170 Figure 66 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 28 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 171 Figure 70 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 29 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 172 Figure 73 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 30 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 173: Replacing The Cassette Unlocking Unit

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 31 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 174 Figure 77 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 32 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 175 Figure 81 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 33 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 176: Replacing The Chip Reader

    Figure 83 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 34 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 177 Figure 85 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 35 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 178 Figure 89 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 36 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 179 Mount in reverse order. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 37 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 180: Replacing The Drawer Unit

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 38 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 181 Mount in reverse order. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 39 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 182: Replacing The Transport Unit Ball Bearings

    30 min. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 40 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 183 For mounting of the ball bearings it is not required to secure them via Loctite again. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 41 03-2009 Type 5175 / 100/110...
  • Page 184: Replacing The Input Rollers

    When ordering, the actual revision of the spare part is delivered. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 42 03-2009 Type 5175 / 100/110...
  • Page 185: Replacing The Side Plate Transport Unit (Left Hand Side)

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 43 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 186 Figure 104 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 44 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 187 Figure 108 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 45 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 188 Mount in reverse order. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 46 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 189: Replacing The Drawer Motor (M4)

    Figure 111 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 47 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 190 Figure 113 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 48 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 191 Figure 117 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 49 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 192 Figure 120 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 50 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 193: Replacing The Opener Stepper Motor (M3)

    30 min. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 51 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 194 Figure 124 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 52 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 195 Figure 128 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 53 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 196 Refer to chapter 3.6, Adjustments. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 54 03-2009 Type 5175 / 100/110...
  • Page 197: Replacing The Roller Stepper Motor (M5)

    30 min. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 55 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 198 Figure 134 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 56 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 199 Figure 138 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 57 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 200 Reject the image. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 58 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 201: Replacing The Roller Clamping Stepper Motor (M6)

    Figure 142 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 59 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 202 Figure 146 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 60 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 203 Do not bend IP guide plate at removal. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 61 03-2009 Type 5175 / 100/110...
  • Page 204 Mount in reverse order. Observe IMPORTANT note below: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 62 03-2009 Type 5175 / 100/110...
  • Page 205 Reject the image. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 63 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 206: Replacing The Roller Ip Sensor

    Figure 155 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 64 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 207 The roller IP sensor is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 65 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 208: Replacing The Toothed Belt Stepper Motor Roller Clamping

    Figure 158 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 66 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 209 Figure 162 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 67 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 210 Do not bend IP guide plate at removal. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 68 03-2009 Type 5175 / 100/110...
  • Page 211 Figure 169 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 69 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 212 Figure 173 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 70 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 213 Figure 177 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 71 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 214 Figure 180 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 72 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 215 The toothed belt Stepper Motor Roller Clamping is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 73 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 216: Replacing The Toothed Belt Stepper Motor Drawer

    Figure 183 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 74 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 217 Figure 187 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 75 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 218 Figure 190 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 76 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 219: Replacing The Toothed Belt Stepper Motor Roller

    Figure 192 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 77 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 220 Figure 194 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 78 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 221 Figure 198 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 79 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 222 The Toothed Belt Stepper Motor Roller is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 80 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 223: Replacing The Toothed Belt Stepper Motor Opener

    Figure 201 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 81 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 224 Figure 205 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 82 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 225 The Toothed Belt Stepper Motor Opener is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 83 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 226: Replacing The User Interface Board

    Figure 209 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 84 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 227 Figure 213 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 85 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 228 The user interface board is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 86 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 229: Replacements At Scan Module

    Figure 216 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 87 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 230 Figure 218 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 88 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 231: Replacing The Pmi Board

    When ordering, the actual revision of the spare part is delivered. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 89 03-2009 Type 5175 / 100/110...
  • Page 232: Replacing The Pmt With Light Collector

    45 min. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 90 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 233 Figure 223 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 91 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 234 Figure 227 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 92 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 235: Replacing The Slow Scan Board

    When ordering, the actual revision of the spare part is delivered. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 93 03-2009 Type 5175 / 100/110...
  • Page 236: Replacing The Slow Scan Unit

    60 min. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 94 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 237 Figure 232 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 95 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 238 The Cassette Unit open Safety Switch is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 96 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 239: Replacing The Distributor Board

    Figure 235 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 97 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 240 Figure 239 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 98 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 241 The Distributor Board is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 99 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 242: Replacing The Power Supply

    Remove all cables in the rear of the digitizer before removal of the rear cover DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 100 03-2009 Type 5175 / 100/110...
  • Page 243 Figure 244 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 101 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 244 Figure 248 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 102 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 245 Figure 251 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 103 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 246 0.3 Ohm with mains cable) check the digitizer for loose grounding connections! DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 104 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 247 The power supply is replaced. Result: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 CR 30-X Chapter 3.5 / 105 03-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 249 ► Referenced Documents Document Title DD+DIS150.06E CR 30-X Service Documentation Chapter 3.5, Replacements / Repair Procedures DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3...
  • Page 250 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 251 Stall Calibration ........................35 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 3 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 252: Overview Of Adjustments And Calibrations

    During preventive maintenance DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 4 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 253: Adjustments

    30 minutes DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 5 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 254: Preparing The Drawer Unit Adjustment

    Figure 3 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 6 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 255 Figure 6 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 7 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 256 Figure 8 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 8 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 257: Checking And Adjusting The Height

    Figure 11 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 9 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 258 Figure 13 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 10 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 259 The drawer unit adjustment is performed. Result DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 11 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 260: Pmi Board Carrier Position Adjustment

    Figure 15 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 12 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 261 Figure 18 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 13 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 262: Cassette Unit Safety Switch Position Adjustment

    20 minutes DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 14 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 263: Checking The Switch Position

    Mount covers. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 15 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 264: Cassette Edge Sensor Position Adjustment

    20 minutes DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 16 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 265 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 17 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 266 The cassette edge sensor is adjusted. Result DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 18 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 267: Laser Beam Position Check And Adjustment

    When ordering, the actual revision of the spare part is delivered. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 19 10-2009 Type 5175 / 100/110...
  • Page 268: Performing The Laser Beam Position Check

    2.5.3 Adjustment DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 20 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 269 Figure 26: Functional principle of the laser beam position check and adjustment DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 21 10-2009 Type 5175 / 100/110...
  • Page 270 Figure 29 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 22 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 271 (10) Log-off the current user at the control PC. (11) Login as user “crservice”. (12) Select menu: <Start - Programs - Agfa CR 30-X - Service Client> (13) Enter user name and password. (14) Select menu: <Analysis and Repair – Check laser beam position>: A window "ESF Initialization"...
  • Page 272 Set switch to ON position. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 24 10-2009 Type 5175 / 100/110...
  • Page 273 Figure 35 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 25 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 274 Figure 38 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 26 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 275 Man ufa ctur er AGFA HEALTHCARE N.V. Sep testr aat 27 - 26 40 - Mor tsel - BE L GIUM ME D I C A L E L E C T R I C A L...
  • Page 276: Verifying The Laser Beam Position Check And Adjustment

    The laser beam position check and adjustment is performed. Result DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 28 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 277: Calibrations

    Mount the Cu-filter below the collimation window at the X-ray device. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 29 10-2009 Type 5175 / 100/110...
  • Page 278 After the second exposure the cassette has been exposed in total with a dose of approximately 20 µGy. (11) Remove the Cu-filter from the X-ray device. (12) Start the CR 30-X Service Client: Select “Start - Programs - Agfa CR 30-X - Service Client”. (13) Select "Preventive Maintenance - Calibration": The current shading calibration is displayed.
  • Page 279 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 31 10-2009 Type 5175 / 100/110...
  • Page 280: Verifying A Successful Shading Calibration

    NOTE:  For external partners accessing the “Troubleshooting Guide” on MedNet, GSO Library may be restricted. In such a case please contact your local Agfa representative.  Checking the images may only happen on light boxes and viewing stations that comply with the specified installation and configuration conditions for diagnostic reading.
  • Page 281: Possible Reasons For A Failed Shading Calibration

    Figure 45 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 33 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 282 Figure 48 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 34 10-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 283: Stall Calibration

    After replacement of the IP handling board (as this has the motor controllers built in)  During preventive maintenance REQUIRED TOOLS: CR 30-X service client on processing station REQUIRED TIME: Approximately 1 minute per stall calibration Start the CR 30-X Service Client. Enter username and password.
  • Page 284 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 3.6 / 36 10-2009 Type 5175 / 100/110...
  • Page 285 E-Label. Document Node ID: 12153275 eq_03-7_menus-setting_e_template_v01 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 3.7 / I Type 5175/100 Agfa Company Confidential...
  • Page 287 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 01-2011 printed in Germany Document Node ID: 20471940 eq_03-8_release-patch_e_template_v06 Agfa Company Confidential Copyright © 2011 Agfa HealthCare N.V.
  • Page 288 Improper operation or service activities may cause damage or injuries. INSTRUCTION:  Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment. ...
  • Page 289 New Features of Software CRE_1511..................18 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 3 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 290 New Features of Software CRE_1901..................24 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 4 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 291: Software Release Cre_1010A

    09/2006  1001 Digitizer SNs Availability MEDNET, GSO Library* (8 MB size) Prerequisites None Reference Document DD+DIS365.06E, Service Bulletin No. 01, CR 30-X System released for Sales (Intranet Link Extranet Link) Supported Functions All functions as specified:  Open the inserted cassette (15 x 30 cm; 18 x 24 cm; 24 x 30 cm; 35 x 43 cm) and transport the IP to the Scan Unit.
  • Page 292: Software Patch Cre_1206 With Patch Cre_1206A

     Solved several items which led to lost communication CR30-X - NX workstation. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 6 01-2011 Type 5175/100/105/110/120/125...
  • Page 293: Software Patch Cre_1206C

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 7 01-2011 Type 5175/100/105/110/120/125...
  • Page 294: Software Patch Cre_1206D

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 8 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 295: Software Release Cre_1309

    Digitizer SNs 1745, 1750, 1753, 1770, 1779, 1780, 1786, 1790, 1793, 1795, 1799, 1800, 1802, 1804, 1809, 1812, 1813, 1814  Part of service kit CM+9 5175 9130 0 (See CR 30-X Availability Service Bulletin No. 12  MEDNET, GSO Library* (8 MB size)
  • Page 296: Software Patch Cre_1309A

    CRE_1206X to CRE_1309 in the field. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 10 01-2011 Type 5175/100/105/110/120/125...
  • Page 297: Software Release Cre_1405

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 11 01-2011 Type 5175/100/105/110/120/125...
  • Page 298: Solved Bugs With Software Cre_1405 Compared To Cre_1309A

    2000 SAL. Expected: 1800 SAL. In software CRE_1405 the reference between IP code and sensitivity is optimized.  CR 30-X Service Client Function "Diagnose Cycle" can be executed, it stays however with screen "Loading ESF". It is not possible to check any function by the diagnose cycle.
  • Page 299: New Features Of Software Cre_1509

    Technical Documentation, chapter 3.2, Tools. Figure 1 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 13 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 300: Solved Bugs With Software Cre_1509 Compared To Cre_1405

    The NX processing station is always switched on. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 14 01-2011 Type 5175/100/105/110/120/125...
  • Page 301: After Software Upgrade Image Plate Transport Stops Before Scan Unit Due To Corrupted Slow Scan Board E-Label (Hq_0708220004)

    A preliminary solution for this symptom was already published with digitizer SB 20, DD+DIS329.07E. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 15 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 302: Additional, Minor Bug Fixes

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 16 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 303: Open Issues

    On MedNet, GSO Library always the latest available software release can be downloaded. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 17 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 304: New Features Of Software Cre_1511

    CRE_1607A. See next section for details. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 18 01-2011 Type 5175/100/105/110/120/125...
  • Page 305: Software Release Cre_1607A

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 19 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 306: Solved Bugs With Software Cre_1607 Compared To Cre_1509 Or Cre_1511

    E-Label data wrong due to corrupted E-label data or wrong service actions. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 20 01-2011 Type 5175/100/105/110/120/125...
  • Page 307: Error 12800 After Switch-On Of The Digitizer (Hq_0901140001)

    PC) despite of successful image transmission. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 21 01-2011 Type 5175/100/105/110/120/125...
  • Page 308: Solved Bug In Software Cre_1607A Compared To Software Cre_1607

    In software CRE_1607A the timing for the boot up of the PMI board is adapted. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 22 01-2011 Type 5175/100/105/110/120/125...
  • Page 309: Software Release Cre_1703

     Improved error messages DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 23 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 310: Software Release Cre_1901

    Improved stall calibration DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 24 01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 311 For Error 10760 the user text is now displayed. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 4 CR 30-X Chapter 3.8 / 25 01-2011 Type 5175/100/105/110/120/125...
  • Page 312 This page is intentionally left blank to enable print out of even and odd pages on duplex printers.
  • Page 313 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 07/2008 printed in Germany Document Node ID: 12152421 eq_03-9_faq_e_template_v05 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 314 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 3.9 / 2 07/2008 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 315 General Schematics Sheet 1 to 9 Document Node ID: 12077164 eq_04_diagrams_e_template_v01 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 4 / I Type 5175/100 Agfa Company Confidential...
  • Page 317: Reference Diagrams

    Slow Scan Stepper Motor (M9) Figure 2, page 4 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 4 / 1 Type 5175/100 Agfa Company Confidential...
  • Page 318: Printed Circuit Boards

    Interconnector XK3 Figure 3, page 5 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 4 / 2 CR30-X Edition 1, Revison 0 Agfa Company Confidential Type 5175/100...
  • Page 319 IP Handling Board 517504aa.cdr User Interface Figure 1 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 4 / 3 Type 5175/100 Agfa Company Confidential...
  • Page 320 517504ab.cdr Stepper Motor (M5) Figure 2 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 4 / 4 CR30-X Edition 1, Revison 0 Agfa Company Confidential Type 5175/100...
  • Page 321 Slow-Scan Lift Position Sensor Figure 3 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 4 / 5 Type 5175/100 Agfa Company Confidential...
  • Page 323: Overview Of Circuit Diagrams

    Wire Harness 5175.8585 Sheet 9 of 9 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 4 / 7 Type 5175/100 Agfa Company Confidential...
  • Page 324 +24VDC_ PMI_POWER_DISTR: +24VDC_ PMI_POWER_DISTR: -W924,-W925,-W926,-W927 #13127 RD,RD,RD,RD /Light_Collector.1A +24VDC_CASSETTE_UNIT -W932,-W933,-W934,-W935 #13126 RD,RD,RD,RD /Cassette_Unit.3A +24VDC_SCANNER_UNIT -W940,-W941,-W942,-W943 #13128 RD,RD,RD,RD /Scanner_Unit.2A +24VDC_ERASURE_UNIT -W948,-W949,-W950,-W951 #13125 RD,RD,RD,RD /ERASURE_UNIT.2A -XK3 F7.0473.6056.0 -BU92 -BU93 -W924,-W925 -W932,-W933 Cassette Unit open RD,RD RD,RD -BU91 -BU94 Safety Switch -W926,-W927 -W934,-W935 Main-Switch RD,RD RD,RD...
  • Page 325 USER_INTERFACE F8.5175.1580.X CHIP READER RollerIpSensor Erasure F8.9499.6940.X +F7.0486.1401.0 f7-9499-6941-1.dfb Reserve Supply on St10 Supply on BU170 toleranc limit +10% / -5% toleranc limit +/- 5% 4 = +24VDC 1,2 = +24VDC 3 = GND 8,9, = GND -W919 F8.5175.8630.0B -BU1#14 *4/4 E-Label F7.0429.0036.0...
  • Page 326 Mirror Laserdiode -W304 -BU50 -BU51 -ST1 -ST2 F8.5175.2920.0 *16/16 *16/16 *16/16 *16/16 Optical lense F7.0429.0029.0 F7.0471.8395.0 F7.0471.8395.0 F7.0429.0029.0 LD not connected Supply on BU50 Supply on BU130 LED1 tolerance limit +-5% 1, 2, 13 = GND 3,4 = +5V (+-2%) 1, 14 = GND (+24V) 5, 6 = -5V (*-2%) 2 = +24V...
  • Page 327 Reset -P17 +24.0V POL -LD53 S/26 F8.5175.8510.0 -W1071 -BU176 -P17 -F17 -10V PMT F7.5175.8515.0 #25/26 #25/26 1,5VD /Scan_Modul.1F -LD32 IN 24VDC -BU129 -F25 24VDC -F13 V33_PHY#12 -L1#12 -BU1#10 *8/8 *8/8 FPGA Test1 M32 Reserve 1 #13127 /FRAME.1E *4/4 F7.0429.0038.0 /FRAME.4F E-Label connector FPGA Test2 M32 Reserve 2...
  • Page 328 SlowScanliftPosSensor F8.9499.5880.X Distributor Board F8.5175.1560.X -W906 -B413 -BU146 -BU144 f7-9499-5881-0.dfb F8.5175.8550.0 *4/4 *4/4 *10/10 *10/10 /Light_Collector.2F F7.0471.8339.0 L7.0429.0009.0 F7.0429.0008.0 F7.0471.8178.0 -W1063 to PMI_P16 F8.5175.2410.0 -W1062 -W1064 -BU145 -BU143 F8.5175.8560.0 F8.5175.2415.0 *4/4 *4/4 *10/10 *10/10 F7.0471.8339.0 L7.0429.0009.0 F7.0429.0008.0 F7.0471.8178.0 Supply on Bu67 -B421 F7-5175-1561-A.dfb tolerance limit + / - 5%...
  • Page 329 Erasure Unit F8.5175.3990.X LD16 LD17 LD31 +B414 -BU61 F8.5175.8510.0 /FRAME.1E *8/8 *8/8 #13125 /FRAME.3F L7.0471.9093.0 F7.0429.0038.0 from Cassette Unit open LD32 LD47 Safety Switch Supply on BU61 toleranc limit + / - 5% -F1#1 1,2,3,4 = 24VDC 5,6,7,8 = GND F7.0459.0304.0 LD48 LD62...
  • Page 330 SCANNER_UNIT -BU96#1 -W940,-W941 RD,RD -BU67#1 BGR F8.5175.8510.0 -W940 -W941 -BU98#1 -W942 -W943 -W944 -BU105#1 -W945 -W946 -W947 -BU112#1 PMI_POWER_DISTR -BU92#1 -BU129#1 -W924 -W925 -BU91#1 -W926 -W927 -W928 SCANNER_UNIT -BU110#1 -W929 -W930 -W931 -BU111#1 -BU134#1 -BU123#1 -W956 -W957 -W958 ERASURE_UNIT -W959 -BU99#1 -BU61#1 -W948...
  • Page 331 BGR F8.5175.8580.0 -2#1 -BU73#1 -W963 -W961 -BU69#1 -BU90#1 -6#1 -BU71#1 -W964 -W962 -BU70#1 -BU97#1 -1#1 -W965 -W965 -BU104#1 -7#1 -W966 -W966 -BU103#1 DATE NAME Wiring Harness Interlock Project USER 15.02.2006 Leith. CR30-X 5175 APPR. Wire_harness_5175.8580 Sheet name: 39319 29.08.06 Leith. NORM Sheet 38631...
  • Page 333 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Document Node ID: 12557915 Edition 1, Revision 3 01-2009 printed in Germany spare_parts_e_template_v07 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 334 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 335 For Recycling Information please refer to: http://intra.agfanet/cd/ep/ehs.nsf DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 3 01-2009 Type 5175/0100/0110 Agfa Company Confidential...
  • Page 336 PAGES 12 - 13 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 4 01-2009 Type 5175/0100/0110 Agfa Company Confidential...
  • Page 337 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 5 01-2009 Type 5175/0100/0110 Agfa Company Confidential...
  • Page 338 Spare Parts List DD+DIS325.06M 5175_CHAP05_01.CDR COVER DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 6 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 339 * =No wearing part. COVER DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 7 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 340 03,04 5175_CHAP05_02.CDR SCAN UNIT DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 8 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 341 * =No wearing part. SCAN UNIT DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 9 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 342 Spare Parts List DD+DIS325.06M 5175_CHAP05_03.CDR SLOWSCAN DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 10 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 343 * =No wearing part. SLOWSCAN DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 11 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 344 07 (M4) 5175_CHAP05_04.CDR DRAWER DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 12 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 345 * =No wearing part. DRAWER DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 13 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 346 5175_CHAP05_05.CDR DRAWER CASSETTE UNIT DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 14 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 347 * =No wearing part. DRAWER CASSETTE UNIT DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 15 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 348 SMALL PARTS AND TOOLS DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 16 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 349 SMALL PARTS AND TOOLS DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 17 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 350 CASSETTE CR MD 4.0 T DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 18 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 351 CASSETTE CR MD 4.0 T DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 19 01-2009 Agfa Company Confidential TYPE 5175/100/110...
  • Page 352 Spare Part Kit version not applicable DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 3 CR 30-X Chapter 5 / 20 01-2009 Type 5175/0100/0110 Agfa Company Confidential...
  • Page 353 Spare Parts List DD+DIS325.06M Type Overview This spare parts list is valid for the following machine type(s): Specification Device Name Type No. CR 30-X 5175/0100 Digitizer (WhiteLabel) 5175/0110 Accessory Overview Following accesssories are separately available: Order number Accessory TABLE FOR CR30-X...
  • Page 355 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 05-2008 printed in Germany Document Node ID: 24101404 eq_08_modifications_e_template_v07 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 356 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 357 May Modified Calibration Board Service Bulletin # 12, DD+DIS148.07E May Small Parts Kit as part of Digitizer Delivery Spare Parts List CR 30-X, DD+DIS325.06M, Rev. 2 May Introduction of Transport Unit with modified lower Service Bulletin # 12, front Roller DD+DIS148.07E...
  • Page 358 Service Bulletin # 08, DD+DIS397.06E DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 8 / 4 Agfa Company Confidential 05-2008 Type 5175 /100/110...
  • Page 359 Service Bulletin # 08, DD+DIS397.06E Document(s) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 8 / 5 Agfa Company Confidential 05-2008 Type 5175 /100/110...
  • Page 360 Service Bulletin # 12, DD+DIS148.07E Document(s) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 8 / 6 Agfa Company Confidential 05-2008 Type 5175 /100/110...
  • Page 361 Service Bulletin # 12, DD+DIS148.07E Document(s) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 8 / 7 Agfa Company Confidential 05-2008 Type 5175 /100/110...
  • Page 362 DD+DIS150.06E Rev. 1 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 8 / 8 Agfa Company Confidential 05-2008 Type 5175 /100/110...
  • Page 363 Part Small parts kit CM+9 5175 9070 1 Number(s) Referenced Spare Parts List CR 30-X, DD+DIS325.06M, Rev. 2 Document(s) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0...
  • Page 364 1795, 1799, 1800, 1802, 1804, 1809, 1812, 1813, 1814 Part Software CRE_1309 is part of service kit CM+9 5175 9130 0 (See Number(s) CR 30-X Service Bulletin No. 12). It is also available via MEDNET, GSO Library Referenced Service Bulletin # 13, DD+DIS158.07E Document(s) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet.
  • Page 365 Service Bulletin # 20, DD+DIS329.07E Document(s) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 8 / 11 Agfa Company Confidential 05-2008 Type 5175 /100/110...
  • Page 366 Service Bulletin # 23, DD+DIS063.08E Document(s) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 8 / 12 Agfa Company Confidential 05-2008 Type 5175 /100/110...
  • Page 367 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 01-2009 printed in Germany Document Node ID: 12076099 eq_09_maintenance_e_template_v06 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 368 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 369 ► Referenced Documents Document Title DD+DIS150.06E CR 30-X Service Manual DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 3...
  • Page 370 Technical Image Quality Check of the System ...............23 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 4 01-2009 Type 5175 / 100/110...
  • Page 371 MAINTENANCE CHECKLIST ....................31 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 5 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 372: General Information On Preventive Maintenance

    2 hours DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 6 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 373: Required Tools

    When ordering, the actual revision of the spare part is delivered. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 7 01-2009 Type 5175 / 100/110...
  • Page 374: Required Spare Parts

    For details refer to sections 3.1 and 3.4. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 8 01-2009 Type 5175 / 100/110...
  • Page 375: Diagnostics

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 9 01-2009 Type 5175 / 100/110...
  • Page 376: Reset Of The Relative Counters

    Figure 2 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 10 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 377: Checking Current Image Quality

    Figure 5 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 11 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 378 (14) Remove the CU filter from the modality. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 12 01-2009 Type 5175 / 100/110...
  • Page 379 (21) Keep the image for comparison after maintenance. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 13 01-2009 Type 5175 / 100/110...
  • Page 380: Maintenance - Power Off

    (old design) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 14 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 381 Figure 12 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 15 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 382: Cleaning General

    Figure 14 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 16 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 383: Checking The Teflon Tape And Cassette Fixation Steel Band At The Cassette Unit

    DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 17 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 384: Greasing The Spindle At The Cassette Fixation Unit

    Figure 19 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 18 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 385: Cleaning Of The Scan Line

    Figure 20 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 19 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 386: Cleaning Of The Lower Scan Rollers

    Figure 21 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 20 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 387: Checking Of Cassette Condition

    Maintenance – Power on Performing Stall Calibration for all Motors Switch on the digitizer. Start the CR 30-X Service Client. Enter username and password. Select "Preventive Maintenance - Stall Calibration". Select "Calibrate all Parameters". Wait till the stall calibration is finished: Takes approx. 5 minutes.
  • Page 388 Close the cassette unit. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 22 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 389 Figure 25 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 23 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 390 Print the image or send it to the archive (depending on what is available). DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 24 01-2009 Type 5175 / 100/110...
  • Page 391 See also IMPORTANT note on page 27 if a FLFS cassette is used for flatfield evaluation. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 25 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 392 Figure 27: Slow scan direction DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 26 01-2009 Type 5175 / 100/110...
  • Page 393 Figure 28 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 27 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 394 Figure 29 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 28 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 395 Confirm that property "compress" is activated. Select "Backup selected files". In the "save as" dialogue select directory: D:\backup\CR 30-X Save the file with date and serial number in the file name. Copy the file additionally to a USB stick and store the file on a save place.
  • Page 396 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 30 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 397 Image Quality DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 31 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 398: Performing Stall Calibration For All Motors

    OK. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 32 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 399 Date / Signature Customer DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 1 CR 30-X Chapter 9 / 33 01-2009 Type 5175 / 100/110 Agfa Company Confidential...
  • Page 401 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 Document Node ID 24620726 04-2008 printed in Germany service_bulletin_cover_e_template_v01 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 402 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR 30-X Chapter 10 / 2 Agfa Company Confidential 04-2008 Type 5175 /100/110...
  • Page 403 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 11-2010 printed in Germany Document Node ID: 14300929 eq_11_install-planning_e_template_v07 Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
  • Page 404 Improper operation or service activities may cause damage or injuries. INSTRUCTION: Read the "Generic Safety Directions" document (see MEDNET GSO => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
  • Page 405 Updated definition about patient vicinity. See section 10. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 3 11-2010 Type 5175/100/105/110/120/125...
  • Page 406 Dimensions, closed Machine ....................25 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 4 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 407 Dimensions Pallet ........................26 SAFETY STANDARDS......................26 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 5 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 408: Installation Planning-Checklist

    Make yourself familiar with the To be aware of the peculiarities when warnings and notes concerning integrating the digitizer into the CR 30-X system integration. system. Compare the required ambient To prevent installation of the digitizer in...
  • Page 409: Product Description

    DD+DIS172.06E Installation Planning Product Description This is the Installation Planning Document for the CR 30-X digitizers type 5175, subtypes 100, 105, 110, 120 and 125. Name of the digitizers:  Subtypes 100, 105: CR 30-X  Subtypes 110, 120, 125: Digitizer (ex factory no name on the covers) All of the above mentioned subtypes are called "digitizer"...
  • Page 410: Scope Of Delivery And Accessories

     The digitizer is ordered and delivered with a dedicated NX workstation.  It is not possible to connect the digitizer to another type of Agfa CR workstation (e.g. QS 3.5). DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
  • Page 411: Accessories

    See NOTE next page. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 9 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 412 In general it is possible to diagnose and repair the digitizer to a certain extent remotely. To enable remote service of the digitizer remote access has first to be organized. Remote Access is established via the Agfa own Secure Remote Service System (SRSS). For more information to the SRSS see MedNet GSO Library path: Technical Services >...
  • Page 413: Digitizer Dimensions

    100.6 [39.6] Figure 4 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 11 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 414: Dimensions Of Optional Tables

    Dimensions see next page. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 12 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 415 Figure 6: Dimensions of all-in-one cart in cm [inch] DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 13 11-2010 Type 5175/100/105/110/120/125...
  • Page 416: Minimum Free Space For Operation And Maintenance

    This can be performed by the customer. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 14 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 417: System Integration

     If the laptop version of the NX workstation is used: Do not place the laptop on top of the CR 30-X, because the accumulated heat developed by the digitizer and the laptop may affect the operation of the system.
  • Page 418 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 16 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 419: Climatic And Ambient Conditions

    , Appendix B. and Immunity: DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 17 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 420: Transport And Storage Conditions

    Warning: Electrical device. Shock possible. INSTRUCTION:  Do not position Agfa EQUIPMENT so it is difficult to operate the disconnection device when an APPLIANCE COUPLER or separable plug is used as isolation.  Local and International wiring regulations must be observed. Check all supplies and voltages, currents, trips and fuses with the Hospital facilities department or their engineers.
  • Page 421 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 19 11-2010 Type 5175/100/105/110/120/125...
  • Page 422 * concerning digitizer mains cable delivery see IMPORTANT note previous page. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 20 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 423: Electrical Connections At The Digitizer

    Input Connector 517511aj.cdr Figure 11 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 21 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 424: Specifications

    256 MB (one complete image) DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 22 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 425: Electrical Data Digitizer

    65 dB(A) maximum DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 23 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 426: Boot-Up Time And Warm-Up Time

    45%. Slightly reduced image quality DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 24 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 427: Dimensions And Weight

    Digitizer with packing incl. palette (kg / pounds) 106 / 233.7 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 25 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 428 EN ISO 13485: 2003; DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 5 CR 30-X Chapter 11 / 26 11-2010 Type 5175/100/105/110/120/125 Agfa Company Confidential...
  • Page 429  Other equipment that is in contact with the patient is not connected directly to the Agfa system without additional protective measures. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
  • Page 430 Check and discuss all the required measures for the installation by means of this checklist. Remarks on the individual items can be noted on the back of the list. After completion please hand out this checklist to your local Agfa representative Name of Agfa FSE:...
  • Page 431 Glossary ..................1 Document Node ID: 12730189 eq_12_glossary_e_template_v01 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 0 CR30-X Chapter 12 / I Type 5175/100 Agfa Company Confidential...
  • Page 433 1Gy= 0.87 x 10 –2 J/kg Agfa Diagnostic Center Application Entity AE – title ASAP - Package AGFA Scanner Access Package Software that runs on the processing station. It handles all digitizer operations Begin Of Line Begin Of Scan Browser A program that provides an interface to the World Wide Web, e.g.
  • Page 434 Computers must run IP to communicate across the Internet. See also TCP. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 12 / 2 CR30-X Edition 1, Revision 0 Agfa Company Confidential Type 5175/100...
  • Page 435 Leeds Phantom Image Plate to check the technical image quality of the CR. The Leeds Phantom was developed for Agfa by the University of Leeds. Look Up Table milli Röntgen, measure for x-ray dose 1 mR corresponds to 8.7 μGy...
  • Page 436 Window and Leveling of digital images. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Chapter 12 / 4 CR30-X Edition 1, Revision 0 Agfa Company Confidential Type 5175/100...
  • Page 439 Version 1.1 06-2008 Added how to integrate the order list for documentation in a printed manual. See section 1. Edition 1, Revision 2 06-2008 printed in Germany Document Node ID: 16099429 Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
  • Page 440 It is recommended to print these pages separately on a DIN A3 or ANSI B (Ledger) printer. DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 Page 2 of 5 06-2008 Agfa Company Confidential...
  • Page 441 Figure 2: Print dialogue for printing single pages or chapters DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 Page 3 of 5 06-2008 Agfa Company Confidential...
  • Page 442 This allows adding comments, to highlight or underline text and many more text manipulations. Figure 4: Adobe Reader comments toolbar DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 Page 4 of 5 06-2008 Agfa Company Confidential...
  • Page 443 DOCUMENT CONTROL NOTE: The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled. Edition 1, Revision 2 Page 5 of 5 06-2008 Agfa Company Confidential...

Table of Contents