Remove the rear film guide.
(6)
Clean the lower postscan roller with
a lint free cloth lightly soaked with
CR screen cleaner.
Turn the clutch at the slow scan unit
to turn the rollers.
Pushing the clutch approx. 60 times
gives one roller rotation.
(7)
Re-mount the guide plates at the pre-scan and posts scan side.
(8)
Re-mount all covers.
(9)
Connect the mains cable.
3.8

Checking of Cassette Condition

(1)
Check the following items of the cassettes:
- outside condition
- shutter
- obvious damage due to drop of cassette
(2)
Keep damaged or defective cassettes for the customer conversation at the end of
the preventive maintenance.
4
Maintenance – Power on
4.1
Performing Stall Calibration for all Motors
(1)
Switch on the digitizer.
(2)
Start the CR 30-X Service Client.
(3)
Enter username and password.
(4)
Select "Preventive Maintenance - Stall Calibration".
(5)
Select "Calibrate all Parameters".
(6)
Wait till the stall calibration is finished: Takes approx. 5 minutes.
(7)
For software < CRE_1309: Switch off/on the digitizer.
For software ≥ CRE_1309 to switch off/on the digitizer is not required.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
5175R3-5_086.cdr
Figure 22
Chapter 9 / 21
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents