Configuring Digitizer Settings - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
NOTE:
If error 2712 pops up on the control PC:
This error indicates, that the installed
software in the Digitizer and on the Control
PC (ASAP software) are not identical.
This error will be solved with steps
described in section
5

Configuring Digitizer Settings

The purpose of this section is to show the required installation tasks.
The following table lists these tasks:
Task
Fill in site specific data.
Check software version.
Configure Maintenance
Indicator.
*Complete required time per Task Code to be entered in the Agfa repair report.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
5.2.
Required
Time
5 minutes
2 minutes
5 minutes
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
The software version of the digitizer and that
one installed on PC do not match.
Please contact service to finish installation.
(Error 2712)
Figure 10
Importance
Task Code*
mandatory
INS
mandatory
INS
optional
INS
Agfa Company Confidential