Diagnostics; Questioning Of The Customer; Checking Images On The Workstation; Checking The Error Statistics - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS171.06E
2

Diagnostics

2.1

Questioning of the Customer

(1)
Ask the customer for any problem that appeared since the last maintenance.
2.2

Checking Images on the Workstation

(1)
Check the last 20 to 40 images on the workstation, to see if image quality
problems occurred.
2.3

Checking the Error Statistics

(1)
On the NX workstation select:
"Start → Programs → CR30-X
Service Client → Reporting → Error
Statistics"
(2)
Check the occurred errors.
2.4

Info Counter Analysis

In the CR30-X Service Client select "General Device Info → Info Counter".
(1)
(2)
Select node "Entire Device".
(3)
Select tab "Values" and check the different entries.
(4)
Select tab "Counters" and check the different entries.
(5)
Select tab "Lists" and check the different entries.
(6)
Click on the errors displayed in the "Lists" tab to check the timestamp: This
shows, when the error occurred.
(7)
Check, whether one of these errors is sporadic and can be solved via software
upgrade, cleaning or a replacement.