AGFA CR 30-X Service Manual page 259

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(2)
Adjust the height with the
height adjustment screws (see
arrows).
(3)
Press down the drawer unit
upper plate after each
adjustment cycle (before
checking the height again).
(4)
Fasten the four height
adjustment fixation screws.
(5)
Again, check the height
adjustment (min. and max.)
with the drawer adjustment tool.
(6)
Repeat adjustment if required.
(7)
Mount IP guide plate, erasure
unit and front cover.
(8)
Connect the mains cable.
(1)
Switch on the digitizer.
Verification
(2)
Wait till the green light at the user interface lights up
(takes approximately 1 minute).
(3)
Scan one image plate with each cassette format the customer is using.
The drawer unit adjustment is performed.
Result
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 14
Chapter 3.6 / 11
Agfa Company Confidential