Page 1
CR 30-X System User Manual 2385A EN 20060630...
Page 2
Agfa-Gevaert N.V. reserves the right to make changes to this document without prior notice. Agfa-Gevaert N.V., Septestraat 27, B-2640 Mortsel, Belgium. CR 30-X, NX, Agfa and Agfa-Rhombus are trademarks of Agfa-Gevaert N.V., Belgium or one of its affiliates. For more information on Agfa products and Agfa HealthCare products, please visit www.agfa.com.
The CR 30-X System Configuration .............. 7 Main System Configuration ................8 Optional System Components................8 Chapter 2:Getting Started with the CR 30-X System ..... 11 Starting the CR 30-X System ..............12 Basic Workflow ..................13 Stopping the CR 30-X System ..............18 Chapter 3:Advanced Function of the CR 30-X System ....
CR 30-X System User Manual (this document). " CR 30-X User Manual, document 2386. " NX User Manual, document 4420 and NX Key User Manual, document 4421. " CR 30-X Plates and Cassettes User Manual, document 2387. Introduction 2385A EN 20060630...
CR 30-X S YSTEM The CR 30-X System Configuration This section covers the following topics: " Main System Configuration " Optional System Components 2385A EN 20060630 Introduction...
The CR 30-X System has the following configuration: • the CR 30-X digitizer, a digitizer for image plates retaining latent X-ray images. The CR 30-X accepts one cassette containing one image plate at a time. • NX 1.0 or higher. NX is a CR workstation for image acquisition, identification, image processing and image transmission of digitized images received from the digitizer.
Page 9
Plug the other side of the UPS power cord into a power outlet. Plug the CR 30-X digitizer and the NX Workstation into the appropriate UPS output receptacles. In case of power failure, the batteries of the UPS supplies power to the CR 30-X and the NX Workstation. 2385A EN 20060630...
Page 10
CR 30-X S YSTEM Introduction 2385A EN 20060630...
Chapter Getting Started with the CR 30-X System This chapter explains how to perform the first basic actions on the CR 30-X System. This chapter covers the following topics: Starting the CR 30-X System Basic Workflow Stopping the CR 30-X System...
Check whether the UPS is connected to a power outlet. Press and hold the On button approximately one second until you hear the UPS beep. Note: Step 1 is only applicable if your CR 30-X System is extended with an uninterruptible power supply (UPS).
CR 30-X S YSTEM Basic Workflow The main functions of the CR 30-X System are digitizing image plates and transmitting the digital image data to the image processing station, where you can perform an image quality control. The workflow consists of the following successive steps: "...
Page 14
Exam (2). • To enter patient data manually, click New Exam (3) and enter patient data and image data manually. For more information, refer to the NX User Manual, document 4420. Getting Started with the CR 30-X System 2385A EN 20060630...
Page 15
Make sure that the cassette is firmly pushed to the right side of the slot [2]. Otherwise, the CR 30-X cannot read the image plate. Note: The cassette is unidentified, so the CR 30-X will send a request to the NX Station. The NX Software must be operational, otherwise the CR 30-X is locked and the status indicator is flashing red.
Page 16
In the Examination window, select the thumbnail in the Image overview pane and click on ID to send the data to the digitizer. As soon as the CR 30-X digitizer has received the complete identification data from the NX Station (via FireWire) it will start digitizing the image plate.
Page 17
Remove the cassette from the cassette slot. Note: When the CR 30-X unlock the cassette, it is ready to be re-used immediately. However, if you leave it for more than 3 days before re-using it, you must re-erase it first.
CR 30-X. Press and hold the Off button until the long beep ceases (approximately five seconds). Note: Step 3 is only applicable if your CR 30-X System is extended with an uninterruptible power supply (UPS). Getting Started with the CR 30-X System...
Page 19
Chapter Advanced Function of the CR 30-X System This chapter provides basic information on how to digitize image plates in emergency situations and how to read the initialization data stored in the memory chip of the cassette tray. Reading an Emergency Image Plate...
In emergency situations, it is possible to open an emergency exam at the NX Station without patient details and to digitize the image plate without having identified the cassette. For detailed information about the emergency license, refer to the NX manuals. Advanced Function of the CR 30-X System 2385A EN 20060630...
To read the initialization data: Check that the CR 30-X System is ready for operation: The status indicator on the CR 30-X digitizer constantly lights up in green. 2385A EN 20060630 Advanced Function of the CR 30-X System...
Page 22
For more information, refer to the NX Key User Manual, document 4421. Click on the Read button at the NX Station. The CR 30-X digitizer waits for the cassette and the status indicator constantly lights up in green. Insert the cassette containing the image plate into the cassette slot [1] of the CR 30-X as shown below.
Page 23
Make sure that the cassette is firmly pushed up to the right side of the slot [2]. Otherwise, the CR 30-X can not read the image plate. Once the cassette is locked, the status indicator on the CR 30-X digitizer is flashing green.
Page 24
CR 30-X S YSTEM Advanced Function of the CR 30-X System 2385A EN 20060630...
Page 28
N.V. reserves the right to make changes to this document without prior notice. Agfa-Gevaert N.V., Septestraat 27, B-2640 Mortsel, Belgium. CR 30-X, NX, Agfa and Agfa-Rhombus are trademarks of Agfa-Gevaert N.V., Belgium or one of its affiliates. Powerware is a trademark of Powerware Corporation.
Page 29
Patient data security ..................20 Maintenance ....................20 Environmental Protection................23 Safety Directions ................... 24 Chapter 2: Getting Started with the CR 30-X ........ 27 Basic Features .................... 28 CR 30-X Features ................... 28 Operating Modes................... 29 The User Interface ..................30 Starting CR 30-X..................
Page 30
CR 30-X Troubleshooting Checklist ................40 Removing a Jammed Image Plate..............43 Behavior in Case of Power Failure ..............45 Appendix A: Equipment Information Sheet ........47 Specifications .................... 48 Appendix B: Remarks for HF-emission and immunity ....51 Remarks for HF-emission and immunity ........... 52...
" Scope This manual contains information for safe and effective operation of the CR 30-X digitizer. Warnings, Cautions, Instructions and Notes The following samples illustrate how warnings, cautions, instructions and notes appear in this document. The text explains their intended use.
Disclaimer Agfa assumes no liability for use of this document if any unauthorized changes have been made to the content or format. Every care has been taken to ensure the accuracy of the information in this document.
CR 30-X Introduction to CR 30-X This section covers the following topics: Intended Use " Intended User " Configuration " System Documentation " Training " Product Complaints " Compatibility " Compliance " Connectivity " Installation " Labels " Patient data security "...
Intended User This manual has been written for trained users of Agfa products and trained diagnostic X–Ray clinical personnel who have received proper training. Users are those persons who actually handle the equipment and those who have authority over the equipment.
CR 30-X Configuration The CR 30-X is always dedicated to a single workstation, on which the identification software as well as the image processing software is running. The identification data are transmitted from the workstation to the digitizer via FireWire. For more information, refer to the workstation’s On-line Help manuals or contact your local service organization.
The documentation shall be kept with the system for easy reference. Technical documentation is available in the product service documentation which is available from your local support organization. Refer to the CR 30-X System User Manual, document 2385, before operating the product. 2386B EN 20060926...
CR 30-X Training The user must have received adequate Agfa training on the safe and effective use of the product before attempting to work with it. Training requirements may vary from country to country. The user should ensure that training is received in accordance with local laws or regulations that have the force of law.
CR 30-X Compatibility The CR 30-X must only be used in combination with other equipment or components if these are expressly recognized by Agfa as compatible. A list of such equipment and components is available from Agfa service on request.
93/42/EEC MDD (European Council Directive 93/42/EEC on Medical Devices). The CR 30-X has been designed in accordance with the IEC 60601-1, Ed. 3: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
Page 41
It is a multilateral agreement among participating countries and certification organizations. Agfa has produced a CB test report and claims national certification in all other member countries of the CB Scheme.
CR 30-X installation and configuration is performed by Agfa. The CR 30-X is equipped with 2 handles at the bottom left and right sides to move the device easily to another location. It is recommended to have at least two persons lift the CR 30-X.
Do not put your fingers in the input slot of the digitizer, they can get hurt when caught between the cassette and the fixation. Insert the cassette as described in the basic workflow of the CR 30-X System User Manual, document 2385.
Page 44
CR 30-X Cassette positioning. Insert the cassette as described in the the basic workflow of the CR 30-X System User Manual, document 2385. Off (power: disconnection from the mains) On (power: connection to the mains) Introduction 2386B EN 20060926...
Page 45
DE S´EXPOSER AU RAYÓN! LASERKLASSE 1 The CR 30-X is a Class 1 Laser Product. It uses one laser diode of a 80 mW type, classification class IIIb. Under normal operating conditions - device with all covers - there can be no laser radiation outside the CR 30-X.
Cleaning the optic unit is required if stripes parallel to the image plate movement can be seen in the image. When you recognize this type of artefact, when using the CR 30-X, clean the optic unit using the cleaning brush. Introduction...
Page 47
CR 30-X To clean the optic unit, proceed as follows: Open the cassette unit. Take out the cleaning brush. Open the lid positioned at the right side of the CR 30-X. Open lid 2386B EN 20060926 Introduction...
Page 48
CR 30-X Clean the scan line. Your last movement must be continuous from the rear to the front. Re-insert the cleaning brush. Close the cassette unit Caution: Misuse of the bowden wire leads to bending, which causes a complicated replacement of the cleaning brush.
CR 30-X Environmental Protection WEEE Notice The Directive on Waste Electrical and Electronic Equipment (WEEE), which came into force as European law on 13th February 2003, resulted in a major change in the treatment of end-of-life electrical equipment. The purpose of...
WARNING: Safety is only guaranteed when trained Agfa personnel has installed the CR 30-X. WARNING: All Agfa medical products must be used by Agfa trained and qualified personnel. WARNING: The user is responsible for judging image quality and controlling environmental conditions for diagnostic softcopy or print viewing.
Page 51
• As is the case for all technical devices, the CR 30-X must be operated, cared for and serviced correctly. A regular quality control is recommended. This can be done according to local standards or by using the Agfa Auto QC tool.
Page 52
(2.29 m*) Patient environment * Deviation for UL 60601-1 • Perform no other operations on the CR 30-X than those described in this document. • Switch the system off before moving it. When reaching the new position, switch the system on again.
Page 53
Chapter Getting Started with the CR 30-X This chapter explains how to perform the first basic actions on the CR 30-X. This chapter covers the following topics: Basic Features Starting CR 30-X Basic Workflow Stopping CR 30-X...
CR 30-X Basic Features CR 30-X Features The CR 30-X is a digitizer for image plates retaining latent X-ray images. It has been developed by Agfa. The CR 30-X accepts one cassette containing one image plate at a time. The "...
CR 30-X Operating Modes CR 30-X can be operated in two modes: operator mode and service mode. Operator Mode The operator mode groups all basic functions which are aimed at radiographers: • Reading an image plate, • Reading an emergency image plate, •...
• the service mode reserved for trained service personnel. The functions of the operator mode are described in this manual. CR 30-X interfaces with the user via: • an erase button, • a status indicator.
Page 57
You can also use the erase button to disable the audio signal after an error. Status Indicator The indicator informs the user by light signals about the status of the CR 30-X. It is positioned at the front of the digitizer, so that it is visible from a distance.
CR 30-X Starting CR 30-X Make sure the CR 30-X is connected to the Control PC and that the Control PC is running the appropriate NX Software. For more information, refer to the NX User Manual. Press the main switch.
Note: The basic workflow is described in the CR 30-X System User Manual, document 2385. Briefly: The CR 30-X starts digitizing the image plate as soon as it has received the complete identification data from the NX Station (via FireWire). 2386B EN 20060926...
CR 30-X Stopping CR 30-X Before Switching Off Check that the CR 30-X is not scanning an image plate. If the CR 30-X is scanning an image plate, the status indicator is flashing green. Switching Off It is recommended to switch off the CR 30-X at the end of the day.
Chapter Operating CR 30-X This chapter provides information about functions that are available in operator mode. Finally you will find some preventive maintenance and troubleshooting guidelines. This chapter covers the following topics: Reading an Emergency Image Plate Re-erasing an Image Plate...
CR 30-X Reading an Emergency Image Plate Note: Reading an emergency image plate is described in the CR 30-X System User Manual, document 2385. Briefly: In emergency situations, it is possible to open an emergency exam at the NX workstation without patient details and to digitize without having identified the cassette.
CR 30-X Re-erasing an Image Plate At the end of a normal or emergency digitizing cycle, the CR 30-X returns an erased image plate. However, in the following cases, you must re-erase the image plate before re-using it in order to prevent ghost images from interfering with the image of interest: •...
Page 64
• The lower part of the status indicator is flashing green. When the CR 30-X has finished erasing the cassette, the upper part of the status indicator is not lit and the lower part is constantly lit in green. Remove the cassette from the cassette slot.
Reading the Initialization Data of an Image Plate Note: Reading the Initialization Data of an Image Plate is described in the CR 30-X System User Manual, document 2385. Briefly: The initialization data stored in the RF-tag of the tray can be read via the CR 30-X.
The only maintenance action which you must perform is checking the image quality. Refer to the User Manual of the NX Software. Troubleshooting Checklist The troubleshooting for a malfunction of the CR 30-X consists of two parts: • The first is always to check the status indicator on the CR 30-X.
Page 67
CR 30-X Errors During Operation If errors occurs during operation, you can consult the Agfa Service Access Package (ASAP) messages on the control PC. The ASAP UI is independent from the NX Software. Status Condition Control PC message Action indicator...
Page 68
“Communication-error. Communication Check if FireWire is Check if fire wire error, FireWire is Red, flashing plugged in or call Agfa connection between not plugged service. digitizer and Control-PC.” Confirm the message Time-out after 5 on the PC and remove “No ID data provided...
The concept provides maximum reliability that no image plate jam can occur in the post scan area. Note: The CR 30-X always reads and digitizes the plate first, then erases it and transports it back into the cassette. If a plate jam occurs before the plate is scanned, there is a fair chance that you can recover the image by putting the image plate back into the cassette and digitizing it again.
Page 70
Note: Never use force to clear the jammed image plate. If it is not possible to gently remove the image plate, call your local service organization. Note: After a jam, the image plate can be used again if not damaged. Close the front cover. Operating CR 30-X 2386B EN 20060926...
Note: The description below is only applicable if an uninterruptible power supply (UPS) is put into the CR 30-X System configuration. For more information about the CR 30-X System, refer to the CR 30-X System User Manual, document 2385. In case of a power failure, the system is still connected to the UPS. Two situations are possible: •...
Page 72
CR 30-X Operating CR 30-X 2386B EN 20060926...
CR 30-X Specifications Product description Type of product Digitizer Commercial name CR 30-X Model number 5175/100 Original seller/manufacturer Agfa-Gevaert NV-Mortsel Labelling 93/42 EEC 'Medical Devices' (Europe), EN 60601-1 UL certified, UL 60601-1 First Edition (North America) cUL certified CSA 22.2 No 601.1...
Page 75
Warming-up time Cold start fully operational after max. 30 min. fully operational after self-test, on condition that the CR 30-X: Warm start • has not been switched off for more than 3 minutes. • has been operational for at least 30 minutes.
Page 76
• 18 x 24 cm: 48 s image plate erase dose. • 15 x 30 cm: 44 s End of Life Estimated product life (if regularly serviced and maintained 7 yrs. according to Agfa instructions) 1. BTU: British Thermal Unit Equipment Information Sheet 2386B EN 20060926...
Page 78
CR 30-X Remarks for HF-emission and immunity This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment. Electromagnetic Environment Transmission Measurements Agreement Guidelines The device uses high frequency energy exclusively for its internal functions.
Page 79
CR 30-X This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment. Resistance to IEC 60601 Level of Electromagnetic Jamming Test Test Level...
Page 80
CR 30-X This device is intended for operation in the electromagnetic environment given below. The user of the device should ensure that it is used in such an environment. Tests of Resistance IEC 60601 Level of Electromagnetic to Disruption Test Level...
Page 81
CR 30-X • REMARK 1: The higher value will apply at 80 MHz and 800 MHz. • REMARK 2: These Guidelines may not apply to all situations. The dispersion of electromagnetic waves is influenced by absorption and reflections from buildings, objects and people.
Page 82
CR 30-X This device is intended for operation in an electromagnetic environment in which the radiated high frequency disturbance variables are monitored. The user of the device can help to prevent electromagnetic disruptions by maintaining the minimum distances between portable and mobile high frequency...
Page 84
0413 Printed in Belgium Published by Agfa-Gevaert N.V., B-2640 Mortsel-Belgium 2386B EN 20060926...
Page 85
CR 30-X Plates and Cassettes User manual 2387A EN 20060913...
Page 86
Agfa-Gevaert N.V. shall under no circumstances be liable for any damage arising from the use or inability to use any information, apparatus, method or process disclosed in this document.
Page 87
CR 30-X Plates and Cassettes Description ......11 Description ....................12 Description of the CR 30-X Cassettes.............. 12 Description of the CR 30-X Tray and Image Plate ........... 14 Initialization ....................15 Initializing Cassette, Tray and Image Plate ............. 15 Replacement Image Plates ................
Page 88
CR 30-X P LATES AND ASSETTES 2387A EN 20060913...
" Scope This manual contains information for safe and effective operation of the CR 30-X Plates and Cassettes. Warnings, Cautions, Instructions and Notes The following samples illustrate how warnings, cautions, instructions and notes appear in this document. The text explains their intended use.
Disclaimer Agfa assumes no liability for use of this document if any unauthorized changes to the content or format have been made. Every care has been taken to ensure the accuracy of the information in this document.
Safety Directions " Intended User This manual has been written for trained users of Agfa products and trained diagnostic X–Ray clinical personnel who have received proper training. Users are those persons who actually handle the equipment and those who have authority over the equipment.
CR 30-X P LATES AND ASSETTES Environmental Protection WEEE Notice The Directive on Waste Electrical and Electronic Equipment (WEEE), which came into force as European law on 13th February 2003, resulted in a major change in the treatment of end-of-life electrical equipment. The purpose of...
ASSETTES Safety Directions General Safety Instructions The CR 30-X tray and image plate have been factory predefined. No further action is required from customer side to use them with the CR 30-X. WARNING: Make sure that the automatic exposure control device is placed above the cassette, to prevent patients from receiving an overdose of X-rays.
Chapter CR 30-X Plates and Cassettes Description This chapter covers the following topics: Description Initialization Precautionary Measures Specifications...
Description Description of the CR 30-X Cassettes The cassettes for the CR 30-X are completely new. The tube side is black. Note: The label on the cassettes illustrated below serves as an example. The name o f the label “CR MD4.0T General” is subject to change.
Page 97
CR 30-X P LATES AND ASSETTES Note: It is not possible to use other cassettes in the CR 30-X. Layout of the CR 30-X cassette: Clip Label Cassette format Lock to open and close the cassette Shutter-opening mechanism 2387A EN 20060913...
CR 30-X P LATES AND ASSETTES Description of the CR 30-X Tray and Image Plate The tray is part of the image plate. The tray supports the image plate and carries the RF tag. Layout of the tray and image plate:...
Initialization Initializing Cassette, Tray and Image Plate For the CR 30-X Plates and Cassettes, the image plate code has been factory embedded in the memory chip (RF-tag) of the tray. This means that the tray and the image plate should always be kept together. A tray and image plate set can be used in any CR 30-X cassette.
CR 30-X P LATES AND ASSETTES Precautionary Measures First Use and Normal Operation • When using new CR plates, you need to manually erase them twice before use. • When CR plates have not been used for 48 hours, you equally need to erase them manually.
CR 30-X P LATES AND ASSETTES Storage • Temperature and humidity levels: Packaged ..........: -25°C to 55°C (-13°F to 131°F) Not packaged .....Temperature..: 15°C to 30°C (59°F to 86°F) ........Relative humidity : 15 to 80% (recommended 30 to 60%) •...
When to clean the tray and image plate? The inner lining of the CR 30-X cassette is made of fleece. This ensures a high degree of protection against electrostatic charging and dust collection on the image plates.
Page 103
CR 30-X P LATES AND ASSETTES How to clean the tray and image plate? Insert a key or a pen into the cassette lock to open the shutter. Remove the key or the pen. Turn the cassette around, so that the black tube side is above.
Page 104
CR 30-X P LATES AND ASSETTES Put the black tray with the image plate onto the cassette, as shown in the figure below. The tray has 2 little hooks. Put the tray onto the cassette in such a way that the hooks reach over the border of the cassette.
Page 105
CR 30-X P LATES AND ASSETTES Put the tray with the image plate back into the cassette. Verify that the white phosphor side is oriented to the black tube side of the cassette and that the shutter does not scratch the image plate.
Page 106
CR 30-X P LATES AND ASSETTES Cleaning the Cassette Clean the cassette thoroughly. Use only a dry cloth or blow away possible dust with compressed air (DO NOT USE A SCREEN CLEANER TO CLEAN THE TRAY.) When necessary, you can clean the outside of the CR MD4.0T cassettes, preferably only with a wet cloth (water, soap).
CR 30-X P LATES AND ASSETTES Specifications Technical Specifications of the CR 30-X Cassette IP Sizes • 35 x 43 cm (14 x 17") • 24 x 30 cm • 18 x 24 cm • 15 x 30 cm Standards •...
Page 108
CR 30-X P LATES AND ASSETTES Identification • Memory chip (RF-tag) embedded in the tray. Note: The tray belongs to the image plate Backscatter Protection • 150 µ lead Environmental Requirements • Temperature: recommended 20-25 °C (allowed: 15-30 °C) •...
CR 30-X P LATES AND ASSETTES Technical Specifications of the CR 30-X Image Plate Sizes • 35 x 43 cm (14 x 17") • 24 x 30 cm • 18 x 24 cm • 15 x 30 cm Plate Construction •...
Page 110
CR 30-X P LATES AND ASSETTES 2387A EN 20060913...
Page 111
CR 30-X P LATES AND ASSETTES 2387A EN 20060913...
Page 112
0413 Printed in Belgium Published by Agfa-Gevaert N.V., B-2640 Mortsel-Belgium 2387A EN 20060913...
Page 113
Dust and dirt build up on CR phosphor plates, leading to image artifacts. To avoid this, plates need cleaning at regular intervals. As CR imaging plates are highly sensitive to water-based cleaning fluids, Agfa has developed an ethanol-based plate cleaner. Using this cleaner, specially designed for CR phosphor plates, will help ensure them the longest service life.
Page 115
X-ray shielding, the CR 30-X can be used inside the X-ray room. Low total cost of ownership Installing the CR 30-X can be done in a single day. With the special LED technology in the erasure unit, no additional electricity is required, so standard electrical outlets are sufficient.
Page 116
CR 30-X digitizer October 2006 CR 30-X is available in combination with NX, Agfa’s image identification and quality control tool, for a highly efficient and optimized radiology workflow. Cassettes with memory CR 30-X uses dedicated cassettes with an embedded memory that stores the data entered during identification.
Page 117
CR 30-X digitizer October 2006 technical s p eCiF iCaTio N s geNeRaL Digitizer type Minimum requirements • Single cassette feed • CR MD 4.0T General Cassette • Throughput: • CR MD 4.0 General Plate 35 x 43 cm (14 x 17”) = approx. 60 plates/hour •...
Page 118
29601 assurance. it certifies that the Quality management system for our X-Ray films conforms to the requirements of annex V of the eeC directive 93/42. agfa healthCare has been certified by lloyd’s Register Canada Quality assurance limited to the following quality management system standards: isO 9001:2000 and en isO 13485-2003.
Page 119
October 2006 COmputed RadiOgRaphy CR MD4.0T General Cassette a UsER-fRiENDly aND DURablE CR CassETTE, foR UsE iN CR 30-X DiGiTizER No change in exposure routine The CR MD4.0T General Cassette is compatible with existing X-ray tables. The exposure equipment and routines do not have to be modified when switching from conventional to digital imaging.