DD+DIS150.06E
ASSEMBLY:
(1)
Mount the cassette opener unit in reverse order.
CALIBRATION:
(1)
Perform a stall calibration for the cassette opener motor.
Refer to chapter 3.6, Adjustments.
(1)
Insert cassette with unexposed IP in the digitizer.
Verification:
(2)
At the workstation select examination type "System diagnosis – Flatfield".
(3)
Select "ID" to identify the image: Scanning starts.
(4)
Confirm that the image arrives at the workstation.
(5)
Reject the image.
(6)
Repeat steps (1) to (5) with each cassette format the customer is using.
IMPORTANT:
In case the boot-up is not successful, check following:
Cable is connected at user interface board.
Cable between chip reader and IP handling board is connected at P 18.
The cassette opener unit is replaced.
Result:
2.6

Replacing the Cassette Unlocking Unit

REQUIRED SPARE PART*:
CM+9 5175 5600 2
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
REQUIRED TOOLS:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.