AGFA CR 30-X Service Manual page 126

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
2.2.16.2
Handling of Replacements in different Software Versions
The following table explains how the different software versions handle the
replacement scenarios described in section 2.2.16.1.
Digitizer Software
Version
≤ CRE_1405
CRE_1509 / 1511
≥ CRE_1607
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
Serial Number of new spare
part = 9999
Module specific data are
automatically written from
Control PC to the board after
first switch on.
Step 1: After exchange and switch on the digitizer detects a
new module with E-Label and issues error 4642.
Step 2: The service engineer enters in the service client,
whether the exchanged part is a submodule of a
larger module (see scenario 1 page) or an own,
adjusted module (see scenario 2 page).
Module specific data are
automatically written from
Control PC to the board after
first switch on.
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Serial Number of new
spare part < 9999
Module specific data are
automatically written from
the board to Control PC
after first switch on.
Step 1: After exchange and
switch on the
digitizer detects a
new module with E-
Label and issues
error 4642.
Step 2: The service
engineer enters in
the service client,
whether the
exchanged part is a
submodule of a
larger module (see
scenario 1 page) or
an own, adjusted
module (see
scenario 2 page).
Agfa Company Confidential