AGFA CR 30-X Service Manual page 258

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
Expected result:
The plate marked with "min."
(= minimum height) has to fit under
the drawer upper plate.
The plate marked with "max."
(= maximum height) may not fit under
the drawer upper plate.
NOTE:
With drawer adjustment tool version 1 it is possible to check the minimum height only.
It is not possible to check the maximum height.
In case the height is OK on both sides there is no height adjustment required.
Evaluation
Otherwise proceed with next steps.
HEIGHT ADJUSTMENT
(1)
Open the four height adjustment
fixation screws (see arrows).
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Min. height
Max. height
Figure 12
517536cd.cdr
Figure 13
Chapter 3.6 / 10