AGFA CR 30-X Service Manual page 174

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
REQUIRED TIME:
60 min.
DISASSEMBLY:
(1)
Switch off the digitizer.
(2)
Disconnect the mains cable.
(3)
Simultaneously press the two buttons (A)
positioned underneath the feed table.
(4)
Pull out cassette unit (B).
(5)
Remove 3 upper screws.
(6)
Open 3 lower screws at front cover a few
turns.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
B
A
Figure 75
Figure 76
5175R3-5_004.cdr
Figure 77
Chapter 3.5 / 32
Agfa Company Confidential
A