DD+DIS150.06E
The drawer unit adjustment is split up in three parts:
#
Drawer Unit Adjustment Part
1
Preparing the adjustment
2

Checking and adjusting the lateral position

3
Checking and adjusting the height position
2.1.1

Preparing the Drawer Unit Adjustment

(1)
Switch off the digitizer.
(2)
Disconnect the mains cable.
(3)
Simultaneously press the two
buttons (A) positioned
underneath the feed table.
(4)
Pull out cassette unit (B).
CAUTION:
Less reliable IP run possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(5)
Remove the front cover.
See chapter 3.5.
(6)
Remove the erasure unit.
See chapter 3.5.
(7)
Slide IP guide plate to the left and
lift it out.