Verifying the Laser Beam Position Check and Adjustment

If only the laser beam position check has been performed:
Verification
(1)
Insert cassette with unexposed IP in the digitizer.
(2)
At the workstation select examination type "System diagnosis – Flatfield".
(3)
Select "ID" to identify the image: Scanning starts.
(4)
Confirm that the image arrives at the workstation.
(5)
Reject the image.
If the laser beam position adjustment has been performed:
Verification is done with the shading calibration procedure.
The laser beam position check and adjustment is performed.
Result