Replacing The Power Supply - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
4.3

Replacing the Power Supply

REQUIRED SPARE PART*:
CM+9 5175 1420 0
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
REQUIRED TOOLS:
REQUIRED TIME:
60 min.
DANGER:
If the power cord is connected and the main switch is switched on the input
side of the power supply is connected to high voltage. Risk of live.
Remove all cables in the rear of the digitizer before removal of the rear cover
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
3 mm
2,5 mm
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Replacements / Repair Procedures
Medium size
Agfa Company Confidential
Figure 241
5 mm
Chapter 3.5 / 100