Replacing The Roller Ip Sensor - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
2.20

Replacing the Roller IP Sensor

REQUIRED SPARE PART*:
CM+9 9499 6950 0
* The last digit in the spare part number indicates the spare
part revision at release of this document. When ordering,
the actual revision of the spare part is delivered.
REQUIRED TOOLS:
None.
REQUIRED TIME:
15 min.
DISASSEMBLY:
(1)
Switch off the digitizer.
(2)
Disconnect the mains cable.
(3)
Simultaneously press the two buttons (A)
positioned underneath the feed table.
(4)
Pull out cassette unit (B).
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 154
B
A
Figure 155
Chapter 3.5 / 64
Agfa Company Confidential
A