AGFA CR 30-X Service Manual page 388

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS171.06E
4.2
Check of Safety Switch
(1)
Pull out the cassette unit approx. 1 cm (approx. 0.4 in.): The "cassette unit open
safety switch" has to switch off the digitizer.
If the test fails, the switch has to be adjusted or it is defective.
For adjustment instructions refer to Technical Documentation chapter 3.6.
For replacement instructions refer to Technical Documentation chapter 3.5.
(2)
Close the cassette unit.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
Figure 23
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Chapter 9 / 22
Agfa Company Confidential