DD+DIS150.06E
(2)
Open the four screws at the optic
module position plates (secured with
red lacquer).
(3)
Shift the optic module gently this
way that both red LEDs (LED 2 and
LED 3) light up.
(4)
Fasten the four screws at the optic
module position plates.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
10-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 36
LED 2
LED 3
LED 1
Switch
Figure 37
Figure 38
Chapter 3.6 / 26