AGFA CR 30-X Service Manual page 205

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
IMPORTANT:
Mount the motor with the motor cable outlet as
shown in the figure enclosed (see arrow):
CALIBRATION:
(1)
Perform a stall calibration for the roller clamping stepper motor.
Refer to chapter 3.6, Adjustments.
(1)
Insert cassette with unexposed IP in the digitizer.
Verification
(2)
At the workstation select examination type "System diagnosis – Flatfield".
(3)
Select "ID" to identify the image: Scanning starts.
(4)
Confirm that the image arrives at the workstation.
(5)
Reject the image.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 153
Chapter 3.5 / 63