DD+DIS150.06E
(3) Select the button <Backup selected files> to perform a backup.
(4) Enter the file name (e.g. backup_CR30-X_SN_XXXX.ZIP) and select the
partition <D:\> of the control PC as storage place.
(5) Wait till the backup is finished (may take up to one minute).
(6) Insert a backup USB stick at the USB port of the control PC.
(7) Copy the backup file stored on the partition <D:\> on the control PC to the
USB stick.
(8) Unmount and remove the USB stick from the control PC and store it on a save
place.
(9)
Close the service client.
(10) Answer the dialogue popping up
"did you perform any repair or
maintenance activities" with "No".
7.2

Hand-over of the System to the Clinical Application Specialist

(1)
Inform the Clinical Application Specialist about:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
11-2009
Controls, Connections, and Setup Procedures
Completion of technical installation.
Any irregularities during the installation.
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
CR 30-X
Figure 25
Chapter 1 / 23
Agfa Company Confidential