DD+DIS150.06E
3

LEDS, Fuses and Switches on the IP Handling Board

T2.5A
F10
5175_chapter3.4_002.cdr
LED
Red LD1
Switch
S1
S2 / 1
S2 / 2
IMPORTANT:
Fuses are displayed only to check fuses in case of a suspected defective fuse.
Always replace the board if a fuse is defective.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
Electrical and mechanical Codes, Fuses, LEDs
T3.5A
T1A
F7
F1#1
Meaning
ON if power for the board available.
Meaning
Board reset. Only for R&D purpose.
ON = normal operation (= default)
OFF = debug position. Only for R&D purpose.
Not used.
DOCUMENT CONTROL NOTE:
Type 5175 / 100/110
T1A
F1#2
Figure 3
CR 30-X
Repair and Service
T500mA
T1A
F8
Chapter 3.4 / 8
Agfa Company Confidential
LD1