Solved Bug In Software Cre_1607A Compared To Software Cre_1607 - AGFA CR 30-X Service Manual

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
The re-transmission of image data failed with "Error 2623" or "Error 2595" when
the connection to the control PC was established again. This happended, when the
FireWire link to the control PC was interrupted and again re-connected.
If the FireWire link to the control PC was interrupted during the image transfer, the
communication between Digitizer and control PC was not established
automatically after re-connection of the FireWire link.
When the cassette was not identified before scanning (warning with error 3138),
the identification on the control PC fails with the error 3096. With software
CRE_1607A the cassette identification on the control PC is now successful.
Errors that appeared in the service client menu "Diagnose Cycle" were correct
displayed only, if the error viewer was restarted.
Software CRE_1607A supports now device serial numbers with 5 digits e.g.
SN10194.
11.3

Solved Bug in Software CRE_1607A compared to Software CRE_1607

Symptom
Digitizer sporadically does not boot-up (HQ_0911270003): After switch-on it stays red
blinking without starting the selftest of motors and sensors. Another switch off/on may
resolve the problem.
Only a few digitizers are affected. At these digitizers the symptom may appear more
frequent.
Cause
Software issue in CRE_1607: The timing during boot-up does not match with certain
hardware parameters of some PMI boards.
Solution
In software CRE_1607A the timing for the boot up of the PMI board is adapted.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
01-2011
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175/100/105/110/120/125
Repair and Service
Software Releases and Patches
Agfa Company Confidential
Chapter 3.8 / 22